FDA Adverse Event Malfunction Summary report: N

QUANTUM MAVERICK BALLOON CATHETER

MDR report key: 1812721 · Received August 24, 2010

Report

Report Number
2134265-2010-03584
Event Type
Malfunction
Date Received
August 24, 2010
Date of Event
July 27, 2010
Report Date
July 28, 2010
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4)

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE, A BALLOON RUPTURE OCCURRED. VASCULAR ACCESS WAS OBTAINED THROUGH THE FEMORAL ARTERY. THE PHYSICIAN WAS PRE-DILATING A 99% STENOSED LESION, APPROXIMATELY 23MM IN LENGTH, LOCATED IN THE LEFT ANTERIOR DESCENDING (LAD) ARTERY WITH THIS 3.0X20MM QUANTUM MAVERICK BALLOON CATHETER. THE BALLOON RUPTURED ON THE FIRST INFLATION AT 13ATMS AFTER BEING INFLATED FOR APPROXIMATELY 7 SECONDS. THE DEVICE WAS REMOVED FROM THE PATIENT INTACT. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUANTUM MAVERICK BALLOON CATHETER CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493808020300 12854450

Patients

Seq Age Sex Outcome Treatment
1 INFLATION DEVICE: MEDTRONIC