NURSE CALL
Report
- Report Number
- 2027454-2023-00095
- Event Type
- Malfunction
- Date Received
- November 13, 2023
- Date of Event
- October 15, 2023
- Report Date
- November 13, 2023
- Manufacturer
- HILL-ROM CARY
- Product Code
- ILQ
- UDI-DI
- 00887761985209
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE NAVICARE® NURSE CALL¿ (NNC) SYSTEM IS A COMPREHENSIVE COMMUNICATION AND INFORMATION MANAGEMENT SYSTEM THAT HELPS CAREGIVERS PROVIDE THE BEST POSSIBLE CARE TO PATIENTS IN THE MOST EFFICIENT MANNER. THE SYSTEM FACILITATES COMMUNICATION WITHIN A NURSING UNIT. CAREGIVERS HAVE THE ABILITY TO QUICKLY LOCATE AND COMMUNICATE WITH A FELLOW STAFF MEMBER, AND PATIENTS CAN EASILY COMMUNICATE WITH CAREGIVERS, THE NURSING STATION, OR DIRECTLY TO THEIR PRIMARY CAREGIVER. DURING TROUBLESHOOTING ACTIVITIES BY BAXTER, IT WAS FOUND THAT THE PBX CONSOLE WAS ONLY MONITORING TWO ROOMS IN THE UNIT. THE CUSTOMER RECALLED THE CVICU UNIT HAD BEEN UNDER CONSTRUCTION FOR TWO MONTHS THIS YEAR AND BELIEVED THERE WAS A MISCOMMUNICATION ON THE HOSPITAL'S PART WHEN THE ROOMS WENT BACK UP TO HAVE THE PBX RE-MONITOR THE OTHER ROOMS. THE BAXTER TECHNICIAN CONFIGURED THE ROOM TO BE MONITORED BY THE PBX CONSOLE AND REQUESTED THE CUSTOMER TEST THE ROOM TO CONFIRM FUNCTIONALITY. IN THIS EVENT, THERE WAS NO PATIENT INJURY, BUT A DELAY IN THE ARRIVAL OF EMERGENCY TEAM MEMBERS OCCURRED BECAUSE THE CODE BLUE CALL DID NOT ROUTE TO THE PBX STATION. A MISSED CODE BLUE CALL BY THE HOSPITAL PBX OPERATOR IS LIKELY TO CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR DEATH IF IT WERE TO RECUR, AS A MANUAL OR AUTOMATED FACILITY-WIDE ANNOUNCEMENT OF THE EVENT AND A SUMMONS FOR ASSISTANCE TO THE CODE BLUE CALL IS COMMONLY PERFORMED BY A PBX STAFF MEMBER. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE COMPLAINT RECORD WILL BE REASSESSED ACCORDINGLY.
THE CUSTOMER REPORTED AN INCIDENT WITH THE NURSE CALL SYSTEM IN WHICH A CODE BLUE CALL FROM ROOM 2036 IN THE CVICU DID NOT ROUTE TO THE PBX STATION. THE CUSTOMER CONFIRMED THE DOME LIGHT OUTSIDE THE PATIENT¿S ROOM ILLUMINATED AND THE CODE BLUE CALL ROUTED TO THE MAIN CONSOLE AT THE NURSE¿S STATION IN THE UNIT. THERE WAS NO PATIENT INJURY REPORTED; HOWEVER, THERE WAS A 5¿10-MINUTE DELAY IN RESUSCITATION EFFORTS AND ARRIVAL OF EMERGENCY TEAM MEMBERS. FOLLOW-UP ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE REPORTED DELAY IN RESUSCITATION EFFORTS WAS UNSUCCESSFUL. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1990896 | NURSE CALL | SYSTEM, COMMUNICATION, POWERED | ILQ | HILL-ROM CARY | 00887761985209 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |