FDA Adverse Event
Malfunction
Summary report: N
GENTLEPOWER LUX CONRA ANGLE 25LPA
MDR report key: 1812460
·
Received August 16, 2010
Report
- Report Number
- 3003637274-2010-00025
- Event Type
- Malfunction
- Date Received
- August 16, 2010
- Date of Event
- July 1, 2010
- Report Date
- July 21, 2010
- Manufacturer
- KAVO DENTAL GMBH
- Product Code
- EFB
- PMA / PMN Number
- K073478
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE USER OF THE DENTAL HANDPIECE RECOGNIZED THE HP IS DEFECTIVE AND GETS HOT. THEREFORE, WAS NOT CAUSED ANY BURN OR OTHER INJURY. DURING PRODUCT EVAL, LOW DEBRIS LEVEL WERE FOUND IN THE HANDPIECE. THE BEARINGS WERE GRITTY AND THE HEAD WAS DAMAGED/DENTED. THE CAUSE FOR OVERHEATING IS THE DENTED HEAD AT THE BACK CAP OF THE HANDPIECE.
Description of Event or Problem · 1
A DENTIST PERFORMED A ROUTINE PROCEDURE ON A PT. DURING TREATMENT, THE DENTAL ELECTRIC HANDPIECE GOT HOT BUT CAUSED NO BURN OR ANY INJURY ON PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENTLEPOWER LUX CONRA ANGLE 25LPA | DENTAL HANDPIECE | EFB | KAVO DENTAL GMBH | 25LPA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |