FDA Adverse Event Injury Summary report: N

CAPTURA HELICAL STONE EXTRACTOR

MDR report key: 1812341 · Received August 12, 2010

Report

Report Number
1825146-2010-00030
Event Type
Injury
Date Received
August 12, 2010
Date of Event
July 16, 2010
Report Date
August 12, 2010
Manufacturer
COOK UROLOGICAL, INC.
Product Code
FFL
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ONE OPENED PACKAGE CONTAINING A USED STONE EXTRACTOR WAS RECEIVED ALONG WITH A SPECIMEN CUP CONTAINING A SMALL CLEAR TUBE. THE CLEAR TUBE SHOULD HAVE MEASURED 1CM IN LENGTH AND WAS NOTED TO BE ACCORDION DOWN TO 5MM. SCRAPE MARKS WERE OBSERVED ON THE BASKET SHEATH LEADING TO THE LOCATION OF THE CLEAR TUBE. THE SHRINK TUBE WAS MOST LIKELY PULLED OFF AND THE BASKET SHEATH WAS DAMAGED WHEN BEING REMOVED ON AN INSTRUMENT OF UNDETERMINED ORIGIN. THERE WAS NO HARM TO THE PATIENT.

Description of Event or Problem · 1

TODAY DURING A CYSTO STONE MANIPULATION CASE, THE DOCTOR INDICATED THE LITTLE PLASTIC PIECE ON THE END OF THE STONE EXTRACTOR POPPED OFF WHEN PULLING THE STONE EXTRACTOR BACK THROUGH THE SCOPE. THE PLASTIC PIECE WAS REMOVED FROM THE PATIENT'S BLADDER USING A GRASPING FORCEP. NUMEROUS COMPLAINTS HAVE BEEN LOGGED FOR THIS SAME ISSUE AND THE DOCTOR NO LONGER WANTS THIS DEVICE AND IS GOING TO USE BOSTON SCIENTIFIC UNTIL COOK FIXES THE PROBLEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPTURA HELICAL STONE EXTRACTOR FFL DISLODGER, STONE, BASKET, URETERAL, METAL FFL COOK UROLOGICAL, INC. NA U1956499

Patients

Seq Age Sex Outcome Treatment
1