FDA Adverse Event Injury Summary report: N

PUMP MMT-723NAS PRDGM INSULIN SMOKE EN

MDR report key: 1812263 · Received August 20, 2010

Report

Report Number
3004209178-2010-82568
Event Type
Injury
Date Received
August 20, 2010
Date of Event
July 19, 2010
Report Date
July 20, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
LZG
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING INCLUDING THE DISPLACEMENT TEST. THE DEVICE WAS PROGRAMMED WITH MULTIPLE BOLUSES AND BASALS AND ALL REGISTERED PROPERLY IN THE BOLUS AND DAILY TOTAL HISTORY. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP WAS GIVING A 25 UNITS OF INSULIN, BUT THE CUSTOMER DENIED. IT WAS STATED THAT THE CUSTOMER CHECKED THE CARELINK AND FOUND THAT THE BOLUS WAS DELIVERED THROUGH THE BOLUS WIZARD. NO FURTHER INFORMATION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-723NAS PRDGM INSULIN SMOKE EN INFUSION PUMP LZG MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 61 YR Hospitalization