FDA Adverse Event Injury Summary report: N

DEXTRUS 4136

MDR report key: 1812191 · Received August 18, 2010

Report

Report Number
1028232-2010-01810
Event Type
Injury
Date Received
August 18, 2010
Date of Event
June 14, 2010
Report Date
July 26, 2010
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVZ
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS RIGHT ATRIAL LEAD WAS ABANDONED SURGICALLY DUE TO DISLODGEMENT. THE LEAD WAS UNABLE TO BE REPOSITIONED DUE TO INGROWN SCAR TISSUE. NO ADVERSE PATIENT EFFECTS WERE REPORTED. A NEW LEAD WAS SUCCESSFULLY IMPLANTED. THIS REPORT WILL BE RE-OPENED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DEXTRUS 4136 PACER LEAD NVZ BIOTRONIK SE & CO. KG 358754

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization