FDA Adverse Event Injury Summary report: N

RSVR MMT-332A 10PK PRDGM 3CC 14L

MDR report key: 1812063 · Received August 18, 2010

Report

Report Number
3004209178-2010-82526
Event Type
Injury
Date Received
August 18, 2010
Date of Event
July 25, 2010
Report Date
July 26, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
FMF
PMA / PMN Number
K032005
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

INSPECTED THREE RANDOM SEALED RESERVOIRS AND THREE OPENED USED RESERVOIRS. MANUAL PRIME TEST AND HIGH-PRESSURE TEST WERE PERFORMED AND THE RESERVOIRS PASSED THE TESTS. NO LEAKS OR DEFECTS IN THE O-RINGS WERE OBSERVED DURING ANALYSIS AND ALL THE RESERVOIRS WERE CONNECTED AND LOCKED IN PLACE PROPERLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED FOR DIABETES KETOACIDOSIS. THE MOTHER STATED THAT SHE CHANGED THE RESERVOIRS SEVERAL TIMES AND THE INSULIN PUMP ALARMED REPEATEDLY NO DELIVERY. NO FURTHER INFO WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RSVR MMT-332A 10PK PRDGM 3CC 14L RESERVOIR FMF MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-332A H7653644

Patients

Seq Age Sex Outcome Treatment
1 16 YR Hospitalization