FDA Adverse Event Malfunction Summary report: N

SPG NEURO STRL XRAY STRG 1X3 200/BX

MDR report key: 18120159 · Received November 13, 2023

Report

Report Number
3011137372-2023-00249
Event Type
Malfunction
Date Received
November 13, 2023
Date of Event
November 6, 2023
Report Date
November 6, 2023
Manufacturer
TELEFLEX MEDICAL
Product Code
EFQ
UDI-DI
04026704417506
PMA / PMN Number
K791871
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4). OTHER REMARKS: N/A. CORRECTED DATA: N/A.

Additional Manufacturer Narrative · 0

(B)(4). ADDITIONAL INFORMATION RECEIVED ON 13 NOV 2023 STATES THE DEFECT WAS FOUND "WHEN CUSTOMER WAS PRE-MOISTENING THE SPONGES PRIOR TO USE ON PATIENT". THE SAMPLE WAS RETURNED TO THE MANUFACTURER (CARWILD CORPORATION) FOR INVESTIGATION. CARWILD REPORTS: "THE INVESTIGATION WAS BASED ON THE REVIEW OF DHR, MANUFACTURING PROCEDURES, HISTORICAL COMPLAINT AND TRAINING RECORDS, AND NO FINDINGS WERE FOUND. SAMPLES THAT WERE PROVIDED SHOWED THAT THE PRODUCT WAS NOT MANUFACTURED BY CARWILD CORP. NO ROOT CAUSE CAN BE DETERMINATE. THERE IS NOT ENOUGH INFORMATION IN THE TELEFLEX NOTIFICATION SUMMARY, SUCH AS: BODY PLACE OF USE, SOLUTION USED, TIME OF THE SPONGE INSIDE THE PATIENT AND SAMPLES SENT WERE NOT MANUFACTURED BY CARWILD CORP." TELEFLEX WILL CONTINUE TO MONITOR AND TREND FOR REPORTED OF THIS NATURE. OTHER REMARKS: N/A. CORRECTED DATA: N/A.

Description of Event or Problem · 0

REPORTED AS "NEURO SPONGES FALLING APART". AT THE TIME OF THIS REPORT THE CUSTOMER HAS NOT RETURNED OUR REQUESTS FOR ADDITIONAL INFORMATION. IF ADDITIONAL INFORMATION IS RECEIVED, THE COMPLAINT FILE WILL BE UPDATED.

Description of Event or Problem · 0

REPORTED AS "NEURO SPONGES FALLING APART". AT THE TIME OF THIS REPORT THE CUSTOMER HAS NOT RETURNED OUR REQUESTS FOR ADDITIONAL INFORMATION. IF ADDITIONAL INFORMATION IS RECEIVED, THE COMPLAINT FILE WILL BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2198102 SPG NEURO STRL XRAY STRG 1X3 200/BX GAUZE/SPONGE, INTERNAL EFQ TELEFLEX MEDICAL 22D1297 04026704417506

Patients

Seq Age Sex Outcome Treatment
1 Unknown