FDA Adverse Event Malfunction Summary report: N

AU640 CHEMISTRY ANALYZER

MDR report key: 1811785 · Received August 23, 2010

Report

Report Number
2050012-2010-00588
Event Type
Malfunction
Date Received
August 23, 2010
Date of Event
July 25, 2010
Report Date
August 23, 2010
Manufacturer
BECKMAN COULTER MISHIMA K.K.
Product Code
JJE
PMA / PMN Number
K961274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER WAS PERFORMING QC ON THE INSTRUMENT AND RECEIVED CUVETTE WASH UNIT ERROR; HOWEVER THE CUSTOMER DISREGARDED THE ERROR MESSAGE. THE (B)(4) ONLINE HELP INSTRUCTS THE USER TO REMOVE OBSTRUCTION IF PRESENT AND CONTACT TECH SUPPORT; HOWEVER, THE CUSTOMER RESET THE INSTRUMENT WITH NO ACTION AND PROCEEDED TO RE-RUN QC. THE CUSTOMER IMMEDIATELY NOTICED THE FLAME AND SMOKE AT THE TOP REAR OF THE INSTRUMENT AND CONTACTED CALL CENTER. A BCI FIELD SERVICE ENGINEER (FSE) INVESTIGATED THE FLAME/SMOKE AND CONCLUDED BASED ON HIS ASSESSMENT THAT THE FLAME APPEARED TO HAVE CAUSED BY A LEAK FROM TUBING LEADING TO DI/DETERGENT DISPENSE MANIFOLD, WHICH MAY HAVE SHORTED THE BOARD. FSE ASSESSMENT IS BASED ON THE LOCATION OF THE FLAME. ALTHOUGH LEAK IS A LIKELY CONTRIBUTOR TO THIS EVENT, A CLEAR ROOT CAUSE CAN NOT BE DETERMINED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO SMOKE BEING RELEASED FROM THE AU640 CHEMISTRY ANALYZER. PATIENT RESULT OR CUSTOMER WAS NOT IMPACTED BY THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AU640 CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER MISHIMA K.K. N/A N/A

Patients

Seq Age Sex Outcome Treatment
1