AU640 CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2010-00588
- Event Type
- Malfunction
- Date Received
- August 23, 2010
- Date of Event
- July 25, 2010
- Report Date
- August 23, 2010
- Manufacturer
- BECKMAN COULTER MISHIMA K.K.
- Product Code
- JJE
- PMA / PMN Number
- K961274
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER WAS PERFORMING QC ON THE INSTRUMENT AND RECEIVED CUVETTE WASH UNIT ERROR; HOWEVER THE CUSTOMER DISREGARDED THE ERROR MESSAGE. THE (B)(4) ONLINE HELP INSTRUCTS THE USER TO REMOVE OBSTRUCTION IF PRESENT AND CONTACT TECH SUPPORT; HOWEVER, THE CUSTOMER RESET THE INSTRUMENT WITH NO ACTION AND PROCEEDED TO RE-RUN QC. THE CUSTOMER IMMEDIATELY NOTICED THE FLAME AND SMOKE AT THE TOP REAR OF THE INSTRUMENT AND CONTACTED CALL CENTER. A BCI FIELD SERVICE ENGINEER (FSE) INVESTIGATED THE FLAME/SMOKE AND CONCLUDED BASED ON HIS ASSESSMENT THAT THE FLAME APPEARED TO HAVE CAUSED BY A LEAK FROM TUBING LEADING TO DI/DETERGENT DISPENSE MANIFOLD, WHICH MAY HAVE SHORTED THE BOARD. FSE ASSESSMENT IS BASED ON THE LOCATION OF THE FLAME. ALTHOUGH LEAK IS A LIKELY CONTRIBUTOR TO THIS EVENT, A CLEAR ROOT CAUSE CAN NOT BE DETERMINED.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO SMOKE BEING RELEASED FROM THE AU640 CHEMISTRY ANALYZER. PATIENT RESULT OR CUSTOMER WAS NOT IMPACTED BY THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AU640 CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER MISHIMA K.K. | N/A | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |