UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2010-00581
- Event Type
- Malfunction
- Date Received
- August 23, 2010
- Date of Event
- August 10, 2010
- Report Date
- August 23, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER NOTED THE LEAK COMING FROM THE HYDRO PNEUMATIC AREA ON THE INSTRUMENT AND SYSTEM GENERATING HIGH VACUUM PRESSURE IS LOW ERROR MESSAGE. THE CUSTOMER HAD PREVIOUSLY EXPERIENCED WATER FILTERING SYSTEM LEAK AND BELIEVED THE LEAK WAS THE SAME AND DID NOT WEAR PERSONAL PROTECTIVE EQUIPMENT (PPE). THE CUSTOMER TRACED THE LEAK BACK TO THE WASH CONCENTRATE REAGENT, WHICH WAS BUBBLING OUT OF THE CONTAINER AND SPLASHED HER HAND. HOTLINE ADVISED CUSTOMER TO WEAR PPE AND PROVIDE CUSTOMER WITH MSDS. A BCI FIELD SERVICE ENGINEER (FSE) REPLACED THE LOOSE VALVE.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT WASH CONCENTRATE REAGENT FROM UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER SPILLED ON THE CUSTOMER HAND CAUSING SLIGHT BURN AND RASH DEVELOPMENT. THE CUSTOMER WAS SENT TO ER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER, | JJE | BECKMAN COULTER INC. | N/A | N/A |