FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 1811784 · Received August 23, 2010

Report

Report Number
2050012-2010-00581
Event Type
Malfunction
Date Received
August 23, 2010
Date of Event
August 10, 2010
Report Date
August 23, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER NOTED THE LEAK COMING FROM THE HYDRO PNEUMATIC AREA ON THE INSTRUMENT AND SYSTEM GENERATING HIGH VACUUM PRESSURE IS LOW ERROR MESSAGE. THE CUSTOMER HAD PREVIOUSLY EXPERIENCED WATER FILTERING SYSTEM LEAK AND BELIEVED THE LEAK WAS THE SAME AND DID NOT WEAR PERSONAL PROTECTIVE EQUIPMENT (PPE). THE CUSTOMER TRACED THE LEAK BACK TO THE WASH CONCENTRATE REAGENT, WHICH WAS BUBBLING OUT OF THE CONTAINER AND SPLASHED HER HAND. HOTLINE ADVISED CUSTOMER TO WEAR PPE AND PROVIDE CUSTOMER WITH MSDS. A BCI FIELD SERVICE ENGINEER (FSE) REPLACED THE LOOSE VALVE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT WASH CONCENTRATE REAGENT FROM UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER SPILLED ON THE CUSTOMER HAND CAUSING SLIGHT BURN AND RASH DEVELOPMENT. THE CUSTOMER WAS SENT TO ER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER, JJE BECKMAN COULTER INC. N/A N/A