FLO-GARD 6201 VOLUMETRIC INFUSION PUMP
MDR report key: 1811764
·
Received August 23, 2010
Report
- Report Number
- 6000001-2010-02706
- Event Type
- Malfunction
- Date Received
- August 23, 2010
- Date of Event
- July 1, 2010
- Report Date
- July 8, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A FLO-GARD INFUSION PUMP WITH A CONSTANT FALSE OCCLUSION ALARM WAS NOT CONFIRMED NOR DUPLICATED DURING PRODUCT EVALUATION. THEREFORE, NO ASSIGNABLE CAUSE CAN BE PROVIDED AND NO REPAIR WAS NECESSARY FOR THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
Description of Event or Problem · 1
THE FACILITY REPORTED A FLO-GARD INFUSION PUMP WITH A CONSTANT FALSE OCCLUSION ALARM. THIS CONDITION WAS DISCOVERED DURING PROGRAMMING/SET-UP IN THE FACILITY'S INTENSIVE CARE AREA. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.