FDA Adverse Event Injury Summary report: N

VASSALLO GT

MDR report key: 18117435 · Received November 12, 2023

Report

Report Number
3012223769-2023-00024
Event Type
Injury
Date Received
November 12, 2023
Date of Event
October 25, 2022
Report Date
November 13, 2023
Manufacturer
FILMECC CO., LTD.
Product Code
DQX
UDI-DI
04562102690047
PMA / PMN Number
K203529NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

FILMECC IS CONDUCTING A RETROSPECTIVE REVIEW OF WORLD-WIDE COMPLAINTS RECEIVED AFTER 510(K) CLEARANCE BUT PRIOR TO COMMERCIAL RELEASE IN THE UNITED STATES. THIS MDR IS BEING FILED BASED ON THE OUTCOME OF THAT RETROSPECTIVE REVIEW. A CAPA HAS BEEN OPENED TO MANAGE THE ACTIONS RELATED TO REMEDIATION OF COMPLAINT FILES AND ANY REQUIRED MDR REPORTING. MANUFACTURING SITE: ASAHI INTECC (THAILAND) CO., LTD., PATHUM THANI, THAILAND, REGISTRATION NUMBER: (B)(4) [MANUFACTURING RECORDS] ALL PRODUCTS WERE CONFIRMED TO HAVE PASSED INSPECTION. [LOT HISTORY REVIEW] WE HAVE RECEIVED ONE CASE OF A SIMILAR EVENT IN THE SAME LOT, BUT AS A RESULT OF THE INVESTIGATION, IT WAS DETERMINED THAT THE EVENT WAS ATTRIBUTABLE TO THE PROCEDURE. [COMPLAINT HISTORY REVIEW] COMPLAINTS ABOUT PRODUCTS OF THE SAME STRUCTURE (VGW1423NS1,VGW1430NS1,VGW1423NS3,VGW1430NS3) OVER THE PAST THREE YEARS SHOWED THAT THE NUMBER OF EVENTS OF SEPARATION SMALL AND DID NOT SHOW AN INCREASING TREND. [RETURNED PRODUCT INVESTIGATION] SINCE THE PRODUCT WAS ALREADY DISCARDED IN THE USER FACILITY, THE PRODUCT WAS NOT AVAILABLE FOR THE INVESTIGATION. BASED ON THE INFORMATION OBTAINED, IT WAS PRESUMED THAT THE CORE WIRE MAY HAVE SEPARATED AND THE COIL MAY HAVE BEEN STRETCHED, LEADING TO SEPARATION, DUE TO BENDING AND STRETCHING AT THE SAME POINT CAUSED BY REPEATED PUSHING AND PULLING WHILE THE TIP OF THE PRODUCT WAS TRAPPED IN THE CALCIFIED AREA IN THE PROCESS OF ATTEMPTING TO PASS THROUGH THE LESION, OR DUE TO A CONTINUOUS ROTATIONAL LOAD IN THE SAME DIRECTION. AS A RESULT OF ABOVE INVESTIGATION, ALTHOUGH IT WAS CONCLUDED THAT THIS EVENT WAS NOT ATTRIBUTABLE TO PRODUCT QUALITY BUT TO THE PROCEDURE, WE CANNOT COMPLETELY RULE OUT THE POSSIBILITY THAT FRACTURED FRAGMENTS WERE LEFT BEHIND IN THE PATIENT'S BODY. INSTRUCTIONS FOR USE (IFU) STATES BELOW AND NO CAPA WILL BE TAKEN. [WARNINGS] NEVER PUSH, AUGER, WITHDRAW, OR TORQUE A GUIDE WIRE THAT MEETS RESISTANCE. TORQUING OR PUSHING A GUIDE WIRE AGAINST RESISTANCE MAY CAUSE GUIDE WIRE DAMAGE AND/OR GUIDE WIRE TIP SEPARATION OR DIRECT DAMAGE TO A VESSEL. RESISTANCE MAY BE FELT AND/OR OBSERVED UNDER FLUOROSCOPY BY NOTING ANY BUCKLING OF THE GUIDE WIRE TIP. IF GUIDE WIRE TIP PROLAPSE IS OBSERVED, DO NOT ALLOW THE TIP TO REMAIN IN A PROLAPSED POSITION.[OTHERWISE DAMAGE TO THE GUIDE WIRE MAY OCCUR.] DETERMINE THE CAUSE OF RESISTANCE UNDER FLUOROSCOPY AND TAKE ANY NECESSARY REMEDIAL ACTION. WHEN TORQUING THIS GUIDE WIRE INSIDE THE BLOOD VESSEL, DO NOT TORQUE CONTINUOUSLY IN THE SAME DIRECTION. THIS MAY CAUSE THE GUIDE WIRE TO BECOME DAMAGED OR BREAK APART, CAUSING INJURY TO THE BLOOD VESSEL OR LEAVING FRAGMENTS INSIDE THE VESSEL. WHEN TORQUING THE GUIDE WIRE, ROTATE IT CLOCKWISE AND COUNTERCLOCKWISE ALTERNATELY. DO NOT EXCEED TWO ROTATIONS (720 DEGREES) IN THE SAME DIRECTION. DO NOT PUSH THE GUIDE WIRE MORE THAN NECESSARY TO ADVANCE THE TIP THROUGH THE NARROWED PART OF THE VESSEL. (FOR EXAMPLE, DO NOT PUSH THE GUIDE WIRE WHEN THE DISTAL TIP OF THE GUIDE WIRE IS BENT BY THE FORCE OF MANIPULATION.) AFTER CROSSING THE TARGETED AREA, DO NOT ROUGHLY TWIST, PUSH OR PULL THE GUIDE WIRE. IF THE GUIDE WIRE IS MOVED EXCESSIVELY, IT MAY BE DAMAGED OR BREAK APART, WHICH MAY INJURE THE BLOOD VESSEL OR LEAVE FRAGMENTS INSIDE THE VESSEL. [MALFUNCTION AND ADVERSE EFFECTS] POSSIBLE COMPLICATIONS AND ADVERSE EVENTS OF GUIDE WIRE USE INCLUDE, BUT ARE NOT LIMITED TO: DAMAGE OF GUIDEWIRE (SEPARATION, BREAKAGE, DAMAGE OF COATING).

Description of Event or Problem · 0

IT WAS REPORTED THAT VASSALLO GT .014 NS1 (THE PRODUCT) WAS USED IN A CASE OF A LESION IN THE BELOW KNEE AREA WITH SEVERE CALCIFICATION AND 100% OCCLUSION. THE PRODUCT WAS TRAPPED IN THE CALCIFIED AREA WHILE PASSING THROUGH THE LESION, CAUSING THE COIL PART OF THE PRODUCT TO STRETCH AND TEAR OFF. THERE WERE NO HEALTH HAZARD ON THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
192931 VASSALLO GT PERIPHERAL GUIDE WIRE DQX FILMECC CO., LTD. NA 220531R021 04562102690047

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other