FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 18115879 · Received November 10, 2023

Report

Report Number
2024168-2023-12532
Event Type
Injury
Date Received
November 10, 2023
Date of Event
October 23, 2023
Report Date
January 11, 2024
Manufacturer
ABBOTT VASCULAR
Product Code
MGB
UDI-DI
08717648013089
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ANALYSIS WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS WAS CONFIRMED. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. THE REPORTED DIFFICULTIES AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING. D4: LOT NUMBER UPDATE FROM 3081241 TO 3081841. H4: MANUFACTURING DATE CORRECTED AS A RESULT.

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT. THE ADDITIONAL PERCLOSE PROGLIDE DEVICE WITH REPORTED ISSUE REFERENCED IN B5 IS FILED UNDER A SEPARATE MEDWATCH REPORT NUMBER.

Description of Event or Problem · 0

IT WAS REPORTED THAT A VENOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL VEIN WAS ATTEMPTED WITH TWO PROGLIDE DEVICES AFTER AN ELECTROPHYSIOLOGY ABLATION INTERVENTIONAL PROCEDURE USING AN 8.5F SHEATH. REPORTEDLY, CUFF MISSES [SUTURE RETRIEVAL ISSUE] OCCURRED WITH BOTH DEVICES. A NEW PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
193579 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB ABBOTT VASCULAR 3081841 08717648013089

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention