CELL-DYN SAPPHIRE VENT HEAD ASSEMBLY
Report
- Report Number
- 2919069-2010-00419
- Event Type
- Malfunction
- Date Received
- August 23, 2010
- Report Date
- July 21, 2010
- Manufacturer
- ABBOTT DIAGNOSTICS DIVISION
- Product Code
- GKZ
- PMA / PMN Number
- D019916
- Removal / Correction Number
- 2919069-7/26/10-005-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(4). CONCOMITANT MEDICAL DEVICE: CELL-DYN SAPPHIRE HEMATOLOGY ANALYZER, LIST # 8H00-01, (B)(4). THE CAUSE OF THE CELL-DYN SAPPHIRE VENT HEAD ASSEMBLY ASPIRATION ERROR (UNDETECTED SHORT SAMPLE) WAS DUE TO A DEFECTIVE CELL-DYN VENT NEEDLE FROM THE SUPPLIER. A PRODUCT RECALL INFORMED ABBOTT CUSTOMERS TO DISCARD ANY CELL-DYN SAPPHIRE VENT HEAD ASSEMBLIES AND REPLACE WITH A NEW CELL-DYN SAPPHIRE VENT HEAD ASSEMBLY PROVIDED TO THE CUSTOMER WITH THE RECALL LETTER.
THE CUSTOMER STATED THE CELL-DYN SAPPHIRE ANALYZER NEEDLE WAS NOT IN THE STARTING POSITION. THE CUSTOMER WAS ADVISED TO REPLACE THE CELL-DYN SAPPHIRE VENT NEEDLE AND THE CUSTOMER PERFORMED A TEST RUN AND THE ERROR WAS NOT GENERATED. THE CUSTOMER OBSERVED QUALITY CONTROLS AND PATIENT RESULTS RUNNING AT APPROPRIATE VALUES. THERE WAS NO IMPACT TO PATIENT MANAGEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CELL-DYN SAPPHIRE VENT HEAD ASSEMBLY | VENT NEEDLE USED ON THE CELL-DYN ANALYZER | GKZ | ABBOTT DIAGNOSTICS DIVISION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CELL-DYN SAPPHIRE ANALYZER, LIST # 8H00-01 |