FDA Adverse Event Malfunction Summary report: N

ESPRIT VENTILATOR

MDR report key: 1811538 · Received August 23, 2010

Report

Report Number
2031642-2010-00233
Event Type
Malfunction
Date Received
August 23, 2010
Date of Event
July 23, 2010
Report Date
July 26, 2010
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
CBK
PMA / PMN Number
K981072
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

POWER SUPPLY.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE VENTILATOR ALARMED AND RESTARTED AND WENT VENT INOP UPON RESTART. THE CUSTOMER REPORTED THE VENTILATOR WAS IN USE ON A PATIENT AND THERE WAS NO PATIENT HARM. THE MANUFACTURER'S FIELD SERVICE TECHNICIAN WAS UNABLE TO DUPLICATE THE CUSTOMER REPORTED EVENT, HOWEVER CONFIRMED A DIAGNOSTIC CODE IN THE VENTILATOR LOG HISTORY THAT INDICATED A +24/+-12 VOLT FAILURE AND RESTART OF THE VENTILATOR. THE SERVICE TECHNICIAN REPLACED THE POWER SUPPLY TO ADDRESS THE FINDINGS. PERFORMANCE VERIFICATION TESTING WAS COMPLETED AND TESTS PASSED PER OPERATING SPECIFICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ESPRIT VENTILATOR VENTILATOR, CONTINUOUS CBK RESPIRONICS CALIFORNIA, INC. V1000 NA