FDA Adverse Event
Malfunction
Summary report: N
TRANSTAR STRETCHER
MDR report key: 1811418
·
Received August 17, 2010
Report
- Report Number
- 1824206-2010-09362
- Event Type
- Malfunction
- Date Received
- August 17, 2010
- Date of Event
- July 23, 2010
- Report Date
- July 23, 2010
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FPO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND A MISSING HEX NUT BOLT THAT CONNECTS FROM THE HEX ROD TO THE BRAKE STEER SHAFT. THE TECHNICIAN INSTALLED A NEW HEX NUT AND READJUSTS THE BRAKE PADS ON BOTH HEAD END CASTER TO REPAIR THE STRETCHER.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATES THE HEAD END CASTERS ARE SLIPPING IN BRAKE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRANSTAR STRETCHER | WHEELED STRETCHER | FPO | HILL-ROM, INC. | 8000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |