LINER: MPACT 01.26.2244MHC DOUBLE MOBILITY HC LINER 22.2/DMB
Report
- Report Number
- 3005180920-2023-00863
- Event Type
- Injury
- Date Received
- November 10, 2023
- Date of Event
- October 13, 2023
- Report Date
- November 10, 2023
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LZO
- UDI-DI
- 07630030806803
- PMA / PMN Number
- K131458
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED ON 17 OCTOBER 2023. LOT 2206939: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 11-MAY-2022. EXPIRATION DATE: 2027-04-22. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED CASE DURING THE PERIOD OF REVIEW. OTHER DEVICES INVOLVED: LINER: MPACT 01.32.3641CF DM CONVERTER TIN COATED D/DMB (K211891) LOT 2110381: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 06-DEC-2021. EXPIRATION DATE: 2026-11-23. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED CASE DURING THE PERIOD OF REVIEW. BALL HEADS: COCR 01.25.124 COCR BALL HEAD 12/14 Ø 22 SIZE S -2,5 (K080885) LOT 1901497: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 18-JUN-2019. EXPIRATION DATE: 2024-06-25. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED CASE DURING THE PERIOD OF REVIEW.
THE PATIENT HAD A PRIMARY HIP SURGERY ON (B)(6) 2023. ON (B)(6) 2023, THE PATIENT CAME IN REPORTING PAIN DUE TO A DISLOCATION OF THE HEAD FROM THE LINER THAT THE PATIENT SUSTAINED WHILE DANCING. THE SURGEON REVISED THE HEAD AND LINER AND THE SURGERY WAS COMPLETED SUCCESSFULLY. PRESENTLY, ON (B)(6) 2023, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1939334 | LINER: MPACT 01.26.2244MHC DOUBLE MOBILITY HC LINER 22.2/DMB | HIP DOUBLE MOBILITY HC LINER | LZO | MEDACTA INTERNATIONAL SA | 2206939 | 07630030806803 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Female | Required Intervention |