FDA Adverse Event
Malfunction
Summary report: N
PUMP MMT-722NAS PRDGM INS V2.2 SK EN
MDR report key: 1811049
·
Received August 10, 2010
Report
- Report Number
- 3004209178-2010-82415
- Event Type
- Malfunction
- Date Received
- August 10, 2010
- Date of Event
- July 13, 2010
- Report Date
- July 24, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD A BLANK DISPLAY AS A RESULT OF MOISTURE DAMAGE FOUND ON THE ELECTRONIC ASSEMBLY. FURTHER TESTING WAS NOT PERFORMED DUE TO THE BLANK DISPLAY.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP WAS EXPOSED TO MOISTURE AND THE DEVICE ALARMED. TROUBLESHOOTING WAS PERFORMED AND THE DISPLAY WAS FROZEN. ALSO, THE MOTHER STATED THAT THE INSULIN PUMP HAD CRACKS AT THE BOTTOM LEFT HAND CORNER. NO FURTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-722NAS PRDGM INS V2.2 SK EN | INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 YR |