FDA Adverse Event Malfunction Summary report: N

PUMP MMT-722NAS PRDGM INS V2.2 SK EN

MDR report key: 1811049 · Received August 10, 2010

Report

Report Number
3004209178-2010-82415
Event Type
Malfunction
Date Received
August 10, 2010
Date of Event
July 13, 2010
Report Date
July 24, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD A BLANK DISPLAY AS A RESULT OF MOISTURE DAMAGE FOUND ON THE ELECTRONIC ASSEMBLY. FURTHER TESTING WAS NOT PERFORMED DUE TO THE BLANK DISPLAY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP WAS EXPOSED TO MOISTURE AND THE DEVICE ALARMED. TROUBLESHOOTING WAS PERFORMED AND THE DISPLAY WAS FROZEN. ALSO, THE MOTHER STATED THAT THE INSULIN PUMP HAD CRACKS AT THE BOTTOM LEFT HAND CORNER. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAS PRDGM INS V2.2 SK EN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 20 YR