FDA Adverse Event
Malfunction
Summary report: N
PUMP MMT-722WWL PRDGM INSULIN CL EN ML
MDR report key: 1811005
·
Received August 9, 2010
Report
- Report Number
- 3004209178-2010-82389
- Event Type
- Malfunction
- Date Received
- August 9, 2010
- Date of Event
- May 15, 2010
- Report Date
- July 26, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD A BLANK DISPLAY AS A RESULT OF A BROKEN SOLDER JOINT AND LIFTED TRACES ON THE INTERFACE BOARD.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP HAD A BLANK DISPLAY. THE BATTERY WAS CHANGED AND A BLANK DISPLAY CONTINUED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-722WWL PRDGM INSULIN CL EN ML | INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-722WWL |