FDA Adverse Event Malfunction Summary report: N

PUMP MMT-722WWL PRDGM INSULIN CL EN ML

MDR report key: 1811005 · Received August 9, 2010

Report

Report Number
3004209178-2010-82389
Event Type
Malfunction
Date Received
August 9, 2010
Date of Event
May 15, 2010
Report Date
July 26, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD A BLANK DISPLAY AS A RESULT OF A BROKEN SOLDER JOINT AND LIFTED TRACES ON THE INTERFACE BOARD.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAD A BLANK DISPLAY. THE BATTERY WAS CHANGED AND A BLANK DISPLAY CONTINUED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722WWL PRDGM INSULIN CL EN ML INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722WWL