FDA Adverse Event Malfunction Summary report: N

PUMP MMT-722NAS PRDGM INS V2.2 SK EN

MDR report key: 1810999 · Received August 9, 2010

Report

Report Number
3004209178-2010-82398
Event Type
Malfunction
Date Received
August 9, 2010
Date of Event
July 6, 2010
Report Date
July 29, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD A BLANK DISPLAY AS A RESULT OF MOISTURE DAMAGE FOUND ON THE ELECTRONIC ASSEMBLY.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT HE JUMPED IN POOL WHILE WEARING THE DEVICE, AND WATER UNDER THE DISPLAY WAS OBSERVED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAS PRDGM INS V2.2 SK EN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 46 YR