FDA Adverse Event
Malfunction
Summary report: N
PUMP MMT-722NAB PRDGM INS V2.2 BL EN
MDR report key: 1810981
·
Received August 9, 2010
Report
- Report Number
- 3004209178-2010-82403
- Event Type
- Malfunction
- Date Received
- August 9, 2010
- Date of Event
- July 9, 2010
- Report Date
- July 29, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD A BLANK DISPLAY AS A RESULT OF CORROSION ON THE RADIO FREQUENCY BOARD, INTERFACE BOARD, AND MOTOR HOME SWITCH.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER JUMPED IN THE WATER WHILE WEARING THE INSULIN PUMP AND THE DEVICE HAD A BLANK DISPLAY. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-722NAB PRDGM INS V2.2 BL EN | INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-722NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 18 YR |