FDA Adverse Event Malfunction Summary report: N

PUMP MMT-722NAB PRDGM INS V2.2 BL EN

MDR report key: 1810981 · Received August 9, 2010

Report

Report Number
3004209178-2010-82403
Event Type
Malfunction
Date Received
August 9, 2010
Date of Event
July 9, 2010
Report Date
July 29, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD A BLANK DISPLAY AS A RESULT OF CORROSION ON THE RADIO FREQUENCY BOARD, INTERFACE BOARD, AND MOTOR HOME SWITCH.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER JUMPED IN THE WATER WHILE WEARING THE INSULIN PUMP AND THE DEVICE HAD A BLANK DISPLAY. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAB PRDGM INS V2.2 BL EN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAB

Patients

Seq Age Sex Outcome Treatment
1 18 YR