LIFEVEST WCD 3100 SYSTEM
Report
- Report Number
- 3002158293-2010-00815
- Event Type
- Malfunction
- Date Received
- August 6, 2010
- Date of Event
- May 22, 2010
- Report Date
- August 6, 2010
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY PACK (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM WAS CONFIRMED. THE CAUSE OF THE DISCHARGED BATTERY PACK WAS DEFECTIVE CELLS WITHIN THE BATTERY PACK. ONE OR MORE OF THE CELLS HAD RESIDUE/CORROSION ON THEM. THE SOURCE OF THE RESIDUE/CORROSION HAS NOT BEEN POSITIVELY IDENTIFIED BUT IS LIKELY ELECTROLYTE FROM A LEAKING CELL. THE ROOT CAUSE OF THE LEAKING CELL WAS NOT POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY PACK. THE PATIENT RECEIVED A REPLACEMENT BATTERY PACK.
THE DAUGHTER OF AN (B)(6) OLD MALE PATIENT CALLED IN TO ZOLL LIFECOR CUSTOMER SUPPORT TO REPORT TROUBLE WITH CHARGING THE PATIENT'S BATTERY. THE PATIENT'S DAUGHTER ALSO REPORTED THAT THE BATTERY WOULD NOT POWER ON THE MONITOR. SUPPORT ISSUED THE PATIENT A REPLACEMENT BATTERY PACK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3100 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR |