FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 3100 SYSTEM
MDR report key: 1810931
·
Received August 6, 2010
Report
- Report Number
- 3002158293-2010-00817
- Event Type
- Malfunction
- Date Received
- August 6, 2010
- Date of Event
- June 30, 2010
- Report Date
- August 6, 2010
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY PACK (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BATTERY PACK FAULTS) WAS CONFIRMED. THE CAUSE OF THE BATTERY FAULTS WAS A BROKEN WHITE WIRE ON THE BATTERY PCA BOARD WHERE THE WIRE ATTACHES TO THE BATTERY CELLS. THE ROOT CAUSE OF THE BROKEN WIRE HAS NOT BEEN DETERMINED BUT MAY HAVE BEEN CAUSED BY A SEVERE SHOCK FROM DROPPING THE PACK. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY PACK. THE PT RECEIVED A REPLACEMENT BATTERY PACK.
Description of Event or Problem · 1
A (B)(6) OLD FEMALE PT CONTACTED ZOLL LIFECOR CUSTOMER SUPPORT TO REPORT THAT SHE HAD BEEN RECEIVING ALARMS RECENTLY. A REVIEW OF THE PT'S DOWNLOAD REVEALED SEVERAL BATTERY CHARGER FAULTS. THE PT WAS PROVIDED WITH A REPLACEMENT BATTERY PACK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3100 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |