BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC)
Report
- Report Number
- 2647876-2023-00452
- Event Type
- Malfunction
- Date Received
- November 9, 2023
- Date of Event
- September 7, 2023
- Report Date
- November 23, 2023
- Manufacturer
- BECTON DICKINSON CARIBE LTD.
- Product Code
- MDB
- UDI-DI
- 00382904420215
- PMA / PMN Number
- K123903
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6 INVESTIGATION SUMMARY: CUSTOMER REPORTED A MOLECULAR FALSE POSITIVE RESULT. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH THE BACTEC PRODUCT BASED ON OUR INTERNAL PROCEDURES AND THE INTENDED USE OF THE PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED, AND BATCH HAS BEEN PREVIOUSLY INVESTIGATED FOR THE REPORTED DEFECT. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS. NO CORRECTIVE ACTIONS WERE REQUIRED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS.
H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
REPORT 1 OF 2. IT WAS REPORTED THAT WHILE USING BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC), THERE WERE TWO MOLECULAR FALSE POSITIVE RESULTS FOR C.TROPICALIS. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER REPORTING FALSE MOLECULAR POSITIVE (IDENTIFICATION OF CANDIDA TROPICALIS) WITH USE OF PRODUCTS 442021 LOT 3130613 BACTEC LYTIC/10 ANAER/F AND THE BIOFIRE PLATFORM. 2 VIALS; LOT 3130613; TESTED (B)(6) 2023; BACTEC INSTRUMENT SERIAL (B)(6); BIOFIRE CATALOG: RFIT-ASY-0147, LOT: 2WDK23. PT. 1: GRAM= GRAM POSITIVE COCCI IN CHAINS; CULTURE RESULT = STREPTOCOCCUS ANGINOSUS; BIOFIRE ID=STREPTOCOCCUS SPECIES AND CANDIDA TROPICALIS"
REPORT 1 OF 2. IT WAS REPORTED THAT WHILE USING BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC), THERE WERE TWO MOLECULAR FALSE POSITIVE RESULTS FOR C.TROPICALIS. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER REPORTING FALSE MOLECULAR POSITIVE (IDENTIFICATION OF CANDIDA TROPICALIS) WITH USE OF PRODUCTS 442021 LOT 3130613 BACTEC LYTIC/10 ANAER/F AND THE BIOFIRE PLATFORM. 2 VIALS; LOT 3130613; TESTED 9/7/2023. BACTEC INSTRUMENT SERIAL (B)(6). BIOFIRE CATALOG RFIT-ASY-0147 LOT 2WDK23. PT. 1: GRAM= GRAM POSITIVE COCCI IN CHAINS; CULTURE RESULT = STREPTOCOCCUS ANGINOSUS; BIOFIRE ID=STREPTOCOCCUS SPECIES AND CANDIDA TROPICALIS"
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 103751 | BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) | SYSTEM, BLOOD CULTURING | MDB | BECTON DICKINSON CARIBE LTD. | 3130613 | 00382904420215 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |