FDA Adverse Event Malfunction Summary report: N

CE INTERMATE LV 50, 24 PACK, 50126

MDR report key: 1810707 · Received August 23, 2010

Report

Report Number
6000001-2010-02695
Event Type
Malfunction
Date Received
August 23, 2010
Date of Event
July 30, 2010
Report Date
July 30, 2010
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
FRN
PMA / PMN Number
K910425
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE DEVICE WAS RECEIVED BY BAXTER FOR EVALUATION. A VISUAL EXAMINATION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF A LEAK FROM THE DISTAL LUER END. THE ROOT CAUSE WAS DETERMINED TO BE INSUFFICIENT BONDING. THIS DEVICE IS A SINGLE USE DEVICE AND WAS DISCARDED. BATCH REVIEW DETERMINED THAT NO NONCONFORMANCE REPORTS WERE DOCUMENTED FOR THIS LOT DURING MANUFACTURING. TREND REVIEW DETERMINED THAT BAXTER HAS RECEIVED SIMILAR REPORTS. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). PER THE CUSTOMER, THE DEVICE IS AVAILABLE FOR EVALUATION; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. A FOLLOW-UP REPORT WILL BE SUBMITTED SHOULD THE DEVICE BE RECEIVED AND EVALUATED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER (B)(4) THAT 8 INTERMATE SV50 DEVICES WERE LEAKING FROM THE DISTAL LUER END. THIS IS REPORT NUMBER 5 OF 8. THE DEVICES HAD BEEN FILLED WITH 90 MILLIGRAMS PAMIDRONATE IN 250 MILLILITERS 0.9% SODIUM CHLORIDE WHEN THE LEAKS WERE OBSERVED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Description of Event or Problem · 1

THE LAY USER/PATIENT CONTACTED LIFESCAN ALLEGING THE METER DISPLAYS ERROR 4 MESSAGE. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WERE NO ALLEGATIONS OF HARM OR INJURY. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CE INTERMATE LV 50, 24 PACK, 50126 PUMP, INFUSION FRN BAXTER HEALTHCARE - IRVINE 10B075

Patients

Seq Age Sex Outcome Treatment
1