FDA Adverse Event Malfunction Summary report: N

DA VINCI

MDR report key: 18107034 · Received November 9, 2023

Report

Report Number
2955842-2023-19766
Event Type
Malfunction
Date Received
November 9, 2023
Date of Event
October 20, 2023
Report Date
October 23, 2023
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
GCJ
UDI-DI
00886874116555
PMA / PMN Number
K191736
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE ENDOSCOPE INSTRUMENT WAS ANALYZED AND FOUND TO HAVE NO FAILURE. THE COMPLAINT WAS NOT CONFIRMED BY FAILURE ANALYSIS. ADDITIONAL OBSERVATION, NOT REPORTED BY THE CUSTOMER: CAMERA INSTRUMENT ADAPTER BEARING FRICTION ISSUE WAS NOTED. CORROSION ON IC ELECTRICAL CONTACTS, DISCOLORATION OF HOUSING AND TRANSMITTANCE TEST FAILURE WERE NOTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE CUSTOMER STATED THAT THE UNIVERSAL SURGICAL MANIPULATOR (USM3) STOPPED WORKING. DURING THE TROUBLESHOOTING, THE USER DISABLED THE USM 3 BY THEMSELVES BECAUSE OF A RECOVERABLE FAULT 23003. IT WAS NOTED THAT THE CUSTOMER ENCOUNTERED AN IMAGE ORIENTATION ISSUE WITH THE ENDOSCOPE. AFTER THE REBOOT AND INSTALLING A NEW ENDOSCOPE, THE SYSTEM WORKED AS INTENDED. THE PROCEDURE CONTINUED AS PLANNED. THERE WAS NO REPORT OF PATIENT INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE ENDOSCOPE WAS MOVING WITH NON-INTUITIVE MOVEMENTS; IT WAS MOVING FREELY WITH UNCONTROLLED MOVEMENTS. THE PROBLEM WAS OCCURRING SPORADICALLY. THE CUSTOMER USED A SPARE ENDOSCOPE AND COMPLETED THE CASE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
111224 DA VINCI ENDOSCOPE PLUS GCJ INTUITIVE SURGICAL, INC 470057-08 N/A 00886874116555

Patients

Seq Age Sex Outcome Treatment
1 Unknown DA VINCI INSTRUMENTS AND ACCESSORIES