CE INTERMATE LV 50, 24 PACK, 50126
Report
- Report Number
- 6000001-2010-02694
- Event Type
- Malfunction
- Date Received
- August 23, 2010
- Date of Event
- July 30, 2010
- Report Date
- July 30, 2010
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- FRN
- PMA / PMN Number
- K910425
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: THE DEVICE WAS RECEIVED BY BAXTER FOR EVALUATION. A VISUAL EXAMINATION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF A FROM THE DISTAL LUER END. THE ROOT CAUSE WAS DETERMINED TO BE INSUFFICIENT BONDING. THIS DEVICE IS A SINGLE USE DEVICE AND WAS DISCARDED. BATCH REVIEW DETERMINED THAT NO NONCONFORMANCE REPORTS WERE DOCUMENTED FOR THIS LOT DURING MANUFACTURING. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). PER THE CUSTOMER, THE DEVICE IS AVAILABLE FOR EVALUATION; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. A FOLLOW-UP REPORT WILL BE SUBMITTED SHOULD THE DEVICE BE RECEIVED AND EVALUATED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE LAY USER/PATIENT CONTACTED LIFESCAN ALLEGING THE METER DISPLAYS ERROR 4 MESSAGE. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WERE NO ALLEGATIONS OF HARM OR INJURY. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.
IT WAS REPORTED TO BAXTER (B)(4) THAT 8 INTERMATE SV50 DEVICES WERE LEAKING FROM THE DISTAL LUER END. THIS IS REPORT NUMBER 4 OF 8. THE DEVICES HAD BEEN FILLED WITH 90 MILLIGRAMS PAMIDRONATE IN 250 MILLILITERS 0.9% SODIUM CHLORIDE WHEN THE LEAKS WERE OBSERVED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CE INTERMATE LV 50, 24 PACK, 50126 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - IRVINE | 10B075 |