CE INTERMATE LV 50, 24 PACK, 50126
Report
- Report Number
- 6000001-2010-02691
- Event Type
- Malfunction
- Date Received
- August 23, 2010
- Date of Event
- July 30, 2010
- Report Date
- July 30, 2010
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- FRN
- PMA / PMN Number
- K910425
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4). PER THE CUSTOMER, THE DEVICE IS AVAILABLE FOR EVALUATION; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. A FOLLOW-UP REPORT WILL BE SUBMITTED SHOULD THE DEVICE BE RECEIVED AND EVALUATED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). DEVICE EVALUATION: THE DEVICE WAS RECEIVED BY BAXTER FOR EVALUATION. A VISUAL EXAMINATION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF A FROM THE DISTAL LUER END. THE ROOT CAUSE WAS DETERMINED TO BE INSUFFICIENT BONDING. THIS DEVICE IS A SINGLE USE DEVICE AND WAS DISCARDED. BATCH REVIEW DETERMINED THAT NO NONCONFORMANCE REPORTS WERE DOCUMENTED FOR THIS LOT DURING MANUFACTURING. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
IT WAS REPORTED TO BAXTER (B)(4) THAT 8 INTERMATE SV50 DEVICES WERE LEAKING FROM THE DISTAL LUER END. THIS IS REPORT NUMBER 1 OF 8. THE DEVICES HAD BEEN FILLED WITH 90 MILLIGRAMS PAMIDRONATE IN 250 MILLILITERS 0.9% SODIUM CHLORIDE WHEN THE LEAKS WERE OBSERVED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
THE LAY USER/PATIENT CONTACTED LIFESCAN ALLEGING THE METER DOES NOT POWER ON. THE METER ALSO DISPLAYED UNKNOWN ERROR MESSAGES. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WERE NO ALLEGATIONS OF HARM OR INJURY. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CE INTERMATE LV 50, 24 PACK, 50126 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - IRVINE | 10B075 |