FDA Adverse Event Injury Summary report: N

POLYSITE IMPLANTABLE INFUSION PORTS

MDR report key: 18104804 · Received November 9, 2023

Report

Report Number
2245270-2023-00080
Event Type
Injury
Date Received
November 9, 2023
Date of Event
October 24, 2023
Report Date
February 20, 2024
Manufacturer
PEROUSE MEDICAL
Product Code
LJT
PMA / PMN Number
K122834
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

WE RECEIVED ONE USED SAMPLE FOR EVALUATION. THE DEVICE RECEIVED IS COMPOSED OF THE IMPLANTABLE PORT, THE CATHETER (25.3 CM), AND ITS RING. THE SERIAL NUMBER OF THE RETURNED PORT IS (B)(4). WE MANAGED TO TRACE LOT NUMBER 23010086 USING THE SERIAL LOT. ACCORDING TO THE AVAILABLE INFORMATION AND OUR EXPERIENCE, WE BELIEVE THAT THE MOST LIKELY HYPOTHESIS THAT COULD EXPLAIN THE CATHETER DISCONNECTION FROM THE IMPLANTABLE PORT COULD BE EXCESSIVE PRESSURE IN THE DEVICE, FOLLOWING A PHENOMENON OF OBSTRUCTION, LEADING TO THE DISCONNECTION OF THE CATHETER AND ITS MIGRATION. THE TITANIUM OUTLET TUBE SHOWS SEVERE DAMAGE, (PINCHED AND CRACKED). WE ASSUME THAT THIS DAMAGE WAS GENERATED DURING THE REMOVAL OF THE PORT, AS OTHERWISE, THE SURGEON WOULD NOT HAVE BEEN ABLE TO CONNECT THE CATHETER. DIMENSIONAL TESTS HAVE BEEN PERFORMED ON THE CATHETER AND THE RING: THE RESULTS ARE COMPLIANT WITH SPECIFICATIONS. FUNCTIONAL TESTS PERFORMED ON THE CATHETER AND THE PORT ARE COMPLIANT. HOWEVER, DURING THIS TEST, WE OBSERVED THE PRESENCE OF BLOOD IN SIGNIFICANT QUANTITIES, ESPECIALLY COAGULATED BLOOD IN THE CATHETER. THE RISK OF DAMAGE TO THE CATHETER DURING ATTEMPTING TO UNBLOCK UNDER PRESSURE IS CLEARLY DESCRIBED IN THE INSTRUCTION FOR USE (IFU # CD00632/003) OF POLYSITE (SECTION III-A WARNING AND PRECAUTIONS). EXTRACT: « NEVER TRY TO UNBLOCK THE DEVICE BY PRESSURE AND NEVER USE A SYRINGE SMALLER THAN 10 ML TO AVOID APPLYING EXCESSIVE PRESSURE WITHIN THE DEVICE. EXCESSIVE PRESSURE CAN DAMAGE THE CATHETER OR LEAD TO DISCONNECTION OF THE PORT». FURTHERMORE, TO AVOID ANY OBSTRUCTION OF THE CATHETER BY BLOOD, POSITIVE PRESSURE MUST BE SYSTEMATICALLY APPLIED WHEN REMOVING THE NEEDLE. INDEED, THE ABSENCE OF INJECTION CAUSES BLOOD REFLUX IN THE CATHETER WHICH CAN LEAD TO ITS OBSTRUCTION. THE REVIEW OF THE MANUFACTURING BATCH RECORD OF THE POLYSITE (B)(4) IMPLANTABLE PORT FINAL PRODUCT (LOT 23010086) DID NOT HIGHLIGHT ANY DEVIATION THAT COULD BE THE ROOT CAUSE OF THE INCIDENT. BASED ON VYGON USA COMPLAINT DATABASE (GRAND AVENUE), THIS IS THE FIRST COMPLAINT RECEIVED REGARDING THIS PRODUCT CODE AND LOT NUMBER. CORRECTIVE ACTION: BASED ON THE INVESTIGATION, THE ROOT CAUSE OF THIS COMPLAINT COULD BE EXCESSIVE PRESSURE IN THE DEVICE, LEADING TO THE DISCONNECTION OF THE CATHETER. THEREFORE, NO FURTHER CORRECTIVE ACTION WAS INITIATED BY QUALITY MANAGEMENT.

Additional Manufacturer Narrative · 0

THE MALFUNCTIONING DEVICE WILL BE RETURNED TO VYGON FOR EVALUATION AS PART OF THE COMPLAINT INVESTIGATION. THE RESULTS OF THIS INVESTIGATION ARE PENDING AND WILL BE COMMUNICATED TO FDA WITHIN 30 DAYS OF ITS CONCLUSION VIA FOLLOW UP MDR.

Description of Event or Problem · 0

PORT TUBING BECAME DISLODGED AND MIGRATED DOWN TO PATIENT'S HEART. PATIENT WAS STABLE POST PORT AND TUBING REMOVAL.

Description of Event or Problem · 0

PORT TUBING BECAME DISLODGED AND MIGRATED DOWN TO PATIENT'S HEART. PATIENT WAS STABLE POST PORT AND TUBING REMOVAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
102587 POLYSITE IMPLANTABLE INFUSION PORTS POLYSITE IMPLANTABLE INFUSION PORTS LJT PEROUSE MEDICAL 4018PI 23010086

Patients

Seq Age Sex Outcome Treatment
1 74 YR Male