FDA Adverse Event
Malfunction
Summary report: N
MONOPLOAR CURVED SCISSORS INSTRUMENT
MDR report key: 1810367
·
Received August 20, 2010
Report
- Report Number
- 2955842-2010-00378
- Event Type
- Malfunction
- Date Received
- August 20, 2010
- Date of Event
- July 26, 2010
- Report Date
- July 26, 2010
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, THE TIP OF ONE SCISSOR BLADE IS BROKEN OFF. THE BROKEN BLADE IS APPROXIMATELY 0.160 X 0.420. THE FRACTURE PLANE OF THE BROKEN BLADE IS NOT STRAIGHT AND THE BLADE HAS A CHIP ON THE EDGE BELOW THE MIDPOINT. THE INTACT BLADE DOES NOT EXHIBIT ANY OBVIOUS CRACKING. NO OTHER DAMAGE WAS FOUND.
Description of Event or Problem · 1
IT WAS A REPORTED THAT DURING A DA VINCI SI SURGICAL PROCEDURE, THE TIP OF THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT BROKE OFF AND FELL INTO THE PATIENT. THE BROKEN PIECE WAS RETRIEVED AND THE PLANNED SURGICAL PROCEDURE WAS COMPLETED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MONOPLOAR CURVED SCISSORS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420179-09 | M10091214 662 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM, ACCESSORIES & ESU |