FDA Adverse Event Malfunction Summary report: N

MONOPLOAR CURVED SCISSORS INSTRUMENT

MDR report key: 1810367 · Received August 20, 2010

Report

Report Number
2955842-2010-00378
Event Type
Malfunction
Date Received
August 20, 2010
Date of Event
July 26, 2010
Report Date
July 26, 2010
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, THE TIP OF ONE SCISSOR BLADE IS BROKEN OFF. THE BROKEN BLADE IS APPROXIMATELY 0.160 X 0.420. THE FRACTURE PLANE OF THE BROKEN BLADE IS NOT STRAIGHT AND THE BLADE HAS A CHIP ON THE EDGE BELOW THE MIDPOINT. THE INTACT BLADE DOES NOT EXHIBIT ANY OBVIOUS CRACKING. NO OTHER DAMAGE WAS FOUND.

Description of Event or Problem · 1

IT WAS A REPORTED THAT DURING A DA VINCI SI SURGICAL PROCEDURE, THE TIP OF THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT BROKE OFF AND FELL INTO THE PATIENT. THE BROKEN PIECE WAS RETRIEVED AND THE PLANNED SURGICAL PROCEDURE WAS COMPLETED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MONOPLOAR CURVED SCISSORS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420179-09 M10091214 662