FDA Adverse Event Malfunction Summary report: N

OT PING METER

MDR report key: 1810294 · Received August 20, 2010

Report

Report Number
2939301-2010-07060
Event Type
Malfunction
Date Received
August 20, 2010
Report Date
August 15, 2010
Manufacturer
LIFESCAN, INC.
Product Code
MDS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

THE LAY USER/PATIENT'S PRODUCT(S) HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS:THE METER INVOLVED IN THIS CASE FAILED TESTING. THE METER WAS FOUND TO HAVE CRACKED/BROKEN DISPLAY. IF ANY ADDITIONAL INFORMATION IS AVAILABLE, THE FDA WILL BE NOTIFIED IN A SECOND FOLLOW UP REPORT. AT THIS TIME, WE CONSIDER THIS MATTER CLOSED.

Description of Event or Problem · 1

THE LAY USER/PATIENT CONTACTED LIFESCAN ALLEGING THE METER HAS BLACK MARKS IN THE DISPLAY. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WERE NO ALLEGATIONS OF HARM OR INJURY. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.

Description of Event or Problem · 1

DURING USE OF THE DEVICE FOR A NON-CLINICAL ACTIVITY, THE USER REPORTED AN INTERMITTENT CF08 ERROR. NO OTHER DETAILS REGARDING THE NATURE OF THE EVENT WERE PROVIDED. SINCE THE EVENT OCCURRED DURING A NON-CLINICAL ACTIVITY, THERE WAS NO PT INVOLVEMENT DURING THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT PING METER GLUCOSE MONITORING SYS/KIT MDS LIFESCAN, INC. 3017417

Patients

Seq Age Sex Outcome Treatment
1 12 YR