Narratives
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
THE LAY USER/PATIENT'S PRODUCT(S) HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE METER INVOLVED IN THIS CASE FAILED TESTING. THE METER WAS FOUND TO HAVE CRACKED/BROKEN DISPLAY. IF ANY ADDITIONAL INFORMATION IS AVAILABLE, THE FDA WILL BE NOTIFIED IN A SECOND FOLLOW UP REPORT. AT THIS TIME, WE CONSIDER THIS MATTER CLOSED.
THE LAY USER/PATIENT CONTACTED LIFESCAN ALLEGING THE METER HAS BLACK MARKS IN THE DISPLAY. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WERE NO ALLEGATIONS OF HARM OR INJURY. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 419 MG/DL, 277 MG/DL, AND 111 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.