SPRINT QUATTRO SECURE
MDR report key: 1810112
·
Received August 20, 2010
Report
- Report Number
- 2649622-2010-05472
- Event Type
- Injury
- Date Received
- August 20, 2010
- Date of Event
- March 18, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- ASKU
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THE LEAD IMPEDANCE VARIED FROM 400 TO 1200 OHMS. IT WAS ALSO REPORTED THAT DURING INTERROGATION, THERE WAS NOISE OVERSENSING WHEN THE PATIENT COUGHED. A LEAD FRACTURE WAS SUSPECTED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THE LEAD WAS REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THE LEAD IMPEDANCE VARIED FROM 400 TO 1200 OHMS. IT WAS ALSO REPORTED THAT DURING INTERROGATION, THERE WAS NOISE OVERSENSING WHEN THE PATIENT COUGHED. A LEAD FRACTURE WAS SUSPECTED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Required Intervention |