HARMONIC ACE 23 CM
Report
- Report Number
- 1527736-2007-07529
- Event Type
- Malfunction
- Date Received
- November 12, 2007
- Date of Event
- August 28, 2007
- Report Date
- August 31, 2007
- Manufacturer
- ETHICON ENDO SURGERY, INC. (CINCINNATI)
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). SERIAL NUMBER = NA. EVAL SUMMARY: THE ANALYSIS FOUND THAT THE DEVICE WAS RECEIVED WITH THE BLADE DISCOLORED (BURNT) DUE TO HEAT GENERATED FROM CONTACT BETWEEN THE BLADE AND TISSUE PAD. THE BLADE WAS FOUND TO BE BROKEN AT THE DISCOLORED (BURNT AREA). THIS FAILURE IS CAUSED DUE TO ACTIVATION OF THE DEVICE WHILE CLAMPED WITHOUT TISSUE BEING PRESENT. FOR THIS REASON THE FOLLOWING STATEMENTS WERE INCLUDED IN THE INSTRUCTIONS FOR USE: "CARE SHOULD BE TAKEN NOT TO APPLY PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. THIS CAN RESULT IN DAMAGE TO THE INSTRUMENT." ADDITIONALLY, A BUTTON WAS NOTED TO BE DISLODGED AND INSIDE THE BAG. A POTENTIAL CAUSE OF THIS COULD HAVE BEEN STERILIZATION, DECONTAMINATION AND TRANSFER OF THE INSTRUMENT FROM SHIPPING SITE. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LOBECTOMY THE DEVICE COULD NOT BE USED TO FINISH THE PROCEDURE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC ACE 23 CM | LFL | ETHICON ENDO SURGERY, INC. (CINCINNATI) | NA | D4HA29 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | HANDPIECE| GENERATOR |