FDA Adverse Event Malfunction Summary report: N

LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER 20 MEDIUM-LARGE TITANIUM LIGA

MDR report key: 1810045 · Received November 12, 2007

Report

Report Number
1527736-2007-07527
Event Type
Malfunction
Date Received
November 12, 2007
Report Date
October 1, 2007
Manufacturer
ETHICON ENDO SURGERY, INC (CINCINNATI)
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): YIELDED JAW. EVALUATION SUMMARY: THE ANALYSIS RESULTS FOR THE ER320 INSTRUMENT FOUND THAT IT WAS RETURNED WITH THE JAWS YIELDED. THE INSTRUMENT WAS CYCLED, FED AND FORMED 4 CONFORMING CLIPS AND 1 CLIP WITH GAP. EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED DURING THIS INSPECTION AND TESTING. WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. ALTHOUGH IT IS NOT POSSIBLE TO CONCLUDE HOW THE CIRCUMSTANCES OCCURRED, IT IS KNOWN FROM THE HISTORY OF THE INSTRUMENT THAT WHEN THE JAWS CLOSE ON A HARD OBJECT (NOT ORIGINAL DESIGN INTENT), THE INDUCED ELONGATION IN THE COMPONENT WILL EXCEED THE MATERIAL'S ELASTIC ZONE (COMPRESSION/TENSILE STRESSES INDUCED ON THE JAWS ARE HIGHER THAN ITS YIELD STRENGTH), HENCE THE JAW WIDTH WILL NOT RETURN TO ITS ORIGINAL DIMENSIONS/POSITION AFTER COMPLETION OF FIRE OUT. IN ADDITION AS PER THE INSTRUCTIONS FOR USE "NEVER FIRE THE INSTRUMENT OVER ANOTHER CLIP OR INSTRUMENT. FIRING THE INSTRUMENT IN THIS MANNER MAY DISTORT OR YIELD THE INSTRUMENT JAWS, WHICH CAN CAUSE THE INSTRUMENT TO "SPIT" CLIPS". THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE DID NOT WORK. NO OTHER DETAILS ARE KNOWN. NO PT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER 20 MEDIUM-LARGE TITANIUM LIGA FZP ETHICON ENDO SURGERY, INC (CINCINNATI) NA UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1