FDA Adverse Event Malfunction Summary report: N

BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC)

MDR report key: 18096604 · Received November 8, 2023

Report

Report Number
2647876-2023-00442
Event Type
Malfunction
Date Received
November 8, 2023
Date of Event
September 8, 2023
Report Date
December 7, 2023
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
MDB
UDI-DI
00382904420215
PMA / PMN Number
K123903
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: CATALOG: 442021 BATCH NO.: 3130613. CUSTOMER REPORTED A MOLECULAR FALSE POSITIVE RESULT. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH THE BACTEC PRODUCT BASED ON OUR INTERNAL PROCEDURES AND THE INTENDED USE OF THE PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED, AND BATCH HAS BEEN PREVIOUSLY INVESTIGATED FOR THE REPORTED DEFECT. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS.

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC), THERE WAS ONE MOLECULAR FALSE POSITIVE FOR C.TROPICALIS. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER REPORTING FALSE MOLECULAR POSITIVE (IDENTIFICATION OF CANDIDA TROPICALIS) WITH USE OF PRODUCTS 442021 LOT 3130613 BACTEC LYTIC/10 ANAER/F AND THE BIOFIRE PLATFORM. 1 VIAL; LOT 3130613; TESTED 9/8/2023 BACTEC INSTRUMENT SERIAL (B)(6) BIOFIRE CATALOG RFIT-ASY-0147 LOT 2WDK23. PT. 1: GRAM= GRAM NEGATIVE RODS; CULTURE RESULT= PSEUDOMONAS AERUGINOSA; BIOFIRE ID= PSEUDOMONAS AERUGINOSA AND CANDIDA TROPICALIS."

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC), THERE WAS ONE MOLECULAR FALSE POSITIVE FOR C.TROPICALIS. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER REPORTING FALSE MOLECULAR POSITIVE (IDENTIFICATION OF CANDIDA TROPICALIS) WITH USE OF PRODUCTS 442021 LOT 3130613 BACTEC LYTIC/10 ANAER/F AND THE BIOFIRE PLATFORM. 1 VIAL; LOT 3130613; TESTED 9/8/2023. BACTEC INSTRUMENT SERIAL (B)(6). BIOFIRE CATALOG RFIT-ASY-0147, LOT 2WDK23. PT. 1: GRAM= GRAM NEGATIVE RODS; CULTURE RESULT= PSEUDOMONAS AERUGINOSA; BIOFIRE ID= PSEUDOMONAS AERUGINOSA AND CANDIDA TROPICALIS."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1277443 BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) SYSTEM, BLOOD CULTURING MDB BECTON DICKINSON CARIBE LTD. 3130613 00382904420215

Patients

Seq Age Sex Outcome Treatment
1 Unknown