CARDIOQUIP MODULAR COOLER HEATER
Report
- Report Number
- 3007899424-2023-00420
- Event Type
- Malfunction
- Date Received
- November 8, 2023
- Date of Event
- October 16, 2023
- Report Date
- July 26, 2024
- Manufacturer
- CARDIOQUIP, LLC
- Product Code
- DWC
- PMA / PMN Number
- K102147
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- 003
Narratives
UDI RELATED DATA QUALITY UPDATES ONLY.
FOLLOWING TEST RESULTS OUTSIDE OF CARDIOQUIP SPECIFICATIONS, CARDIOQUIP RECOMMENDED THAT THE DEVICE RECEIVE AN INTERNAL WATER PATHWAY REPLACEMENT. THE CUSTOMER THEN SENT THE DEVICE TO CARDIOQUIP FOR REPAIR. THE DEVICE WAS RETURNED TO SPECIFICATION VIA AN INTERNAL WATER PATHWAY REPLACEMENT. FOLLOWING THE REPAIR, THE DEVICE PASSED INSPECTION AND IS FULLY FUNCTIONAL.
A CUSTOMER REQUESTED DEVICE TESTING DUE TO ALLEGED BACTERIAL CONTAMINATION. HPC TESTING WAS PERFORMED TO PROVIDE A QUANTITATIVE ASSESSMENT OF WATER QUALITY. FOLLOWING TEST RESULTS OUTSIDE OF CARDIOQUIP SPECIFICATIONS, CARDIOQUIP RECOMMENDED THAT THE DEVICE RECEIVE AN INTERNAL WATER PATHWAY REPLACEMENT. THE CUSTOMER AGREED TO THAT SERVICE. CURRENTLY CARDIOQUIP IS WAITING TO RECEIVE THE DEVICE IN ORDER TO PERFORM THE SERVICE. FOLLOWING THE INTERNAL WATER PATHWAY REPLACEMENT, THE DEVICE WILL BE RETURNED TO SPECIFICATION. A FOLLOW-UP WILL BE FILED IF ANY ADDITIONAL INFORMATION OR INFORMATION THAT CORRECTS THIS REPORT IS OBTAINED.
BASED ON THESE RESULTS (10161342 -137000 MPN/ML), THIS DEVICE IS ABOVE THE ACCEPTABLE LIMITS FOR WATER QUALITY IN CARDIOQUIP COOLER-HEATER. TO REMEDIATE THE DEVICE CONTAMINATION, THE FACILITY CAN EITHER PERFORM AN INTERNAL WATER PATH REPLACEMENT (IWPR) OR PARTICIPATE IN CARDIOQUIP'S TRADE-IN PROGRAM.
BASED ON THESE RESULTS (10161342 -137000 MPN/ML), THIS DEVICE IS ABOVE THE ACCEPTABLE LIMITS FOR WATER QUALITY IN CARDIOQUIP COOLER-HEATER. TO REMEDIATE THE DEVICE CONTAMINATION, THE FACILITY CAN EITHER PERFORM AN INTERNAL WATER PATH REPLACEMENT (IWPR) OR PARTICIPATE IN CARDIOQUIP'S TRADE-IN PROGRAM.
BASED ON THESE RESULTS (10161342 -137000 MPN/ML), THIS DEVICE IS ABOVE THE ACCEPTABLE LIMITS FOR WATER QUALITY IN CARDIOQUIP COOLER-HEATER. TO REMEDIATE THE DEVICE CONTAMINATION, THE FACILITY CAN EITHER PERFORM AN INTERNAL WATER PATH REPLACEMENT (IWPR) OR PARTICIPATE IN CARDIOQUIP'S TRADE-IN PROGRAM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1582721 | CARDIOQUIP MODULAR COOLER HEATER | CARDIOPULMONARY BYPASS DEVICE | DWC | CARDIOQUIP, LLC | MCH-1000(I) |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |