FDA Adverse Event Malfunction Summary report: N

CARDIOQUIP MODULAR COOLER HEATER

MDR report key: 18095604 · Received November 8, 2023

Report

Report Number
3007899424-2023-00420
Event Type
Malfunction
Date Received
November 8, 2023
Date of Event
October 16, 2023
Report Date
July 26, 2024
Manufacturer
CARDIOQUIP, LLC
Product Code
DWC
PMA / PMN Number
K102147
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

UDI RELATED DATA QUALITY UPDATES ONLY.

Additional Manufacturer Narrative · 0

FOLLOWING TEST RESULTS OUTSIDE OF CARDIOQUIP SPECIFICATIONS, CARDIOQUIP RECOMMENDED THAT THE DEVICE RECEIVE AN INTERNAL WATER PATHWAY REPLACEMENT. THE CUSTOMER THEN SENT THE DEVICE TO CARDIOQUIP FOR REPAIR. THE DEVICE WAS RETURNED TO SPECIFICATION VIA AN INTERNAL WATER PATHWAY REPLACEMENT. FOLLOWING THE REPAIR, THE DEVICE PASSED INSPECTION AND IS FULLY FUNCTIONAL.

Additional Manufacturer Narrative · 0

A CUSTOMER REQUESTED DEVICE TESTING DUE TO ALLEGED BACTERIAL CONTAMINATION. HPC TESTING WAS PERFORMED TO PROVIDE A QUANTITATIVE ASSESSMENT OF WATER QUALITY. FOLLOWING TEST RESULTS OUTSIDE OF CARDIOQUIP SPECIFICATIONS, CARDIOQUIP RECOMMENDED THAT THE DEVICE RECEIVE AN INTERNAL WATER PATHWAY REPLACEMENT. THE CUSTOMER AGREED TO THAT SERVICE. CURRENTLY CARDIOQUIP IS WAITING TO RECEIVE THE DEVICE IN ORDER TO PERFORM THE SERVICE. FOLLOWING THE INTERNAL WATER PATHWAY REPLACEMENT, THE DEVICE WILL BE RETURNED TO SPECIFICATION. A FOLLOW-UP WILL BE FILED IF ANY ADDITIONAL INFORMATION OR INFORMATION THAT CORRECTS THIS REPORT IS OBTAINED.

Description of Event or Problem · 0

BASED ON THESE RESULTS (10161342 -137000 MPN/ML), THIS DEVICE IS ABOVE THE ACCEPTABLE LIMITS FOR WATER QUALITY IN CARDIOQUIP COOLER-HEATER. TO REMEDIATE THE DEVICE CONTAMINATION, THE FACILITY CAN EITHER PERFORM AN INTERNAL WATER PATH REPLACEMENT (IWPR) OR PARTICIPATE IN CARDIOQUIP'S TRADE-IN PROGRAM.

Description of Event or Problem · 0

BASED ON THESE RESULTS (10161342 -137000 MPN/ML), THIS DEVICE IS ABOVE THE ACCEPTABLE LIMITS FOR WATER QUALITY IN CARDIOQUIP COOLER-HEATER. TO REMEDIATE THE DEVICE CONTAMINATION, THE FACILITY CAN EITHER PERFORM AN INTERNAL WATER PATH REPLACEMENT (IWPR) OR PARTICIPATE IN CARDIOQUIP'S TRADE-IN PROGRAM.

Description of Event or Problem · 0

BASED ON THESE RESULTS (10161342 -137000 MPN/ML), THIS DEVICE IS ABOVE THE ACCEPTABLE LIMITS FOR WATER QUALITY IN CARDIOQUIP COOLER-HEATER. TO REMEDIATE THE DEVICE CONTAMINATION, THE FACILITY CAN EITHER PERFORM AN INTERNAL WATER PATH REPLACEMENT (IWPR) OR PARTICIPATE IN CARDIOQUIP'S TRADE-IN PROGRAM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1582721 CARDIOQUIP MODULAR COOLER HEATER CARDIOPULMONARY BYPASS DEVICE DWC CARDIOQUIP, LLC MCH-1000(I)

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown