FDA Adverse Event Malfunction Summary report: N

TUFTEX OVER-THE-WIRE EMBOLECTOMY CATHETER

MDR report key: 18094317 · Received November 8, 2023

Report

Report Number
1220948-2023-00190
Event Type
Malfunction
Date Received
November 8, 2023
Date of Event
October 5, 2023
Report Date
November 8, 2023
Manufacturer
LEMAITRE VASCULAR, INC.
Product Code
DXE
UDI-DI
00840663100750
PMA / PMN Number
K022145
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE PRODUCT WAS NOT RETURNED FOR EVALUATION. HOWEVER, THE PROVIDED PHOTOS CONFIRMED THE REPORT. WHILE THE REPORTED EVENT WAS CONFIRMED, THE EXACT ROOT CAUSE OF THE BALLOON RUPTURE COULD NOT BE DETERMINED. DUE TO THE NATURE OF THE PRODUCT, BALLOON RUPTURES ARE AN EXPECTED RISK WHEN USING THIS PRODUCT. AS STATED IN THE IFU: AS WITH ALL CATHETERIZATION PROCEDURES, COMPLICATIONS MAY OCCUR. THESE MAY INCLUDE BUT ARE NOT LIMITED TO: INFECTION, LOCAL HEMATOMAS, INTIMAL DISRUPTION, ARTERIAL DISSECTION, PERFORATION AND RUPTURE, HEMORRHAGE, ARTERIAL THROMBOSIS, DISTAL EMBOLI OF BLOOD CLOTS OR ARTERIOSCLEROTIC PLAQUE, AIR EMBOLUS, ANEURYSMS, ARTERIAL SPASMS, ARTERIOVENOUS FISTULA FORMATION, BALLOON RUPTURE OR TIP SEPARATION WITH FRAGMENTATION AND DISTAL EMBOLIZATION. THE POSSIBILITY OF BALLOON RUPTURE MUST BE TAKEN INTO ACCOUNT WHEN CONSIDERING THE RISKS INVOLVED IN THE CATHERIZATION PROCEDURE. THE LOT NUMBER OF THE PRODUCT WAS NOT PROVIDED. THEREFORE, WE'RE UNABLE TO PERFORM A LOT REVIEW. CAPA 2021-009 WAS PREVIOUSLY OPENED TO ADDRESS THIS ISSUE. NO FURTHER ACTION NEEDED AT THIS TIME.

Description of Event or Problem · 0

IT WAS REPORTED TUFTTEX OTW BALLOON RUPTURED DURING A PROCEDURE. NO INJURY WAS REPORTED TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
744015 TUFTEX OVER-THE-WIRE EMBOLECTOMY CATHETER EMBOLECTOMY CATHETER DXE LEMAITRE VASCULAR, INC. XOT1317 00840663100750

Patients

Seq Age Sex Outcome Treatment
1 Unknown