PEDICLE SCREW MUST LT 15MM SHORT - PEDICLE SCREW Ø4.5X30 CANN
Report
- Report Number
- 3005180920-2023-00883
- Event Type
- Malfunction
- Date Received
- November 8, 2023
- Date of Event
- October 12, 2023
- Report Date
- December 11, 2023
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- NKB
- UDI-DI
- 07630971278226
- PMA / PMN Number
- K203482
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- 003
Narratives
DEVICES RECEIVED ON 16 NOBEMBER 2023 AND ANALYZED BY MEDACTA SPINE R&D PROJECT MANAGER ON 21 NOVEMBER 2023: NO VISIBLE DEFECT CAN BE FOUND. THE MOST PROBABLE ROOT CAUSE FOR THE SCREW LOOSENING IS THE WEAKNESS OF THE PATIENT'S BONE.
BATCH REVIEW PERFORMED ON 31-OCT-2023. LOT 2227012: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 08-MAR-2023. EXPIRATION DATE: 2028-02-14. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. 2 ITEMS IN THIS COMPLAINT. PEDICLE SCREW 03.57.003 MUST LT 15MM SHORT - PEDICLE SCREW Ø4.5X35 CANN (K203482) LOT: 2224911: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 13-JUL-2022. EXPIRATION DATE: 2027-06-26. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. 3 ITEMS IN THIS COMPLAINT. PEDICLE SCREW 03.57.003 MUST LT 15MM SHORT - PEDICLE SCREW Ø4.5X35 CANN (K203482) LOT 2222978: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 23-JUN-2022. EXPIRATION DATE: 2027-05-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, NO ITEMS OF THE SAME LOT HAVE BEEN SOLD.
ALL PEDICLE SCREWS LOOSENED DURING FINAL FIXATION OF SET SCREWS (T4-T7 FIXATION). ALL SCREWS WERE REPLACED WITH 30 MINUTES OF DELAY, 4 HOURS OF SURGERY. THE PATIENT BONE WAS WEAK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 813026 | PEDICLE SCREW MUST LT 15MM SHORT - PEDICLE SCREW Ø4.5X30 CANN | PEDICLE SCREW | NKB | MEDACTA INTERNATIONAL SA | 2227012 | 07630971278226 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Male | Other |