FDA Adverse Event Injury Summary report: N

MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL

MDR report key: 18089675 · Received November 8, 2023

Report

Report Number
2032227-2023-298967
Event Type
Injury
Date Received
November 8, 2023
Date of Event
October 21, 2023
Report Date
December 7, 2023
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000545604
PMA / PMN Number
P150001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE PUMP PASSED THE FUNCTIONAL TEST, INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST, AND THE DAT AT 0.0872 INCHES. THE PUMP WAS RECEIVED WITH BROKEN BATTERY TUBE THREADS AND CRACKED CASE AT THE BATTERY TUBE SIDE. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: MINOR SCRATCHED DISPLAY WINDOW, SCRATCHED CASE, AND PILLOWING KEYPAD OVERLAY. THE TEST P-CAP AND RESERVOIR DOES LOCK IN PLACE IN THE RESERVOIR COMPARTMENT. HISTORY DOWNLOAD WAS SUCCESSFUL USING THUMP AND CARELINK UPLOAD WAS SUCCESSFUL. THE PUMP DID NOT HAVE A BATTERY INSTALLED WHEN RECEIVED. THE PUMP WAS RECEIVED WITHOUT THE ORIGINAL BATTERY CAP. PLEASE SEE BELOW FOR THE BOLUSES LISTED ON THE EVENT DATE 21-OCT-2023 IN THE PUMP HISTORY FILE. 10/21/2023 08:27:27.000 NORMALBOLUSDELIVERED (220), BOLUSPROGRAMMINGMETHOD: BOLUSWIZARD (1), NORMALBOLUSAMOUNTPROGRAMMED: 56000 (5.6 U), BOLUSAMOUNTDELIVERED: 56000 (5.6 U). 10/21/2023 12:11:31.000 NORMALBOLUSDELIVERED (220), BOLUSPROGRAMMINGMETHOD: BOLUSWIZARD (1), NORMALBOLUSAMOUNTPROGRAMMED: 78500 (7.85 U), BOLUSAMOUNTDELIVERED: 78500 (7.85 U). 10/21/2023 12:11:31.000 NORMALBOLUSDELIVERED (220), BOLUSPROGRAMMINGMETHOD: BOLUSWIZARD (1), NORMALBOLUSAMOUNTPROGRAMMED: 78500 (7.85 U), BOLUSAMOUNTDELIVERED: 78500 (7.85 U). PLEASE SEE BELOW FOR THE DAILY TOTAL OF ALL INSULIN DELIVERED ON THE EVENT DATE 21-OCT-2023 IN THE PUMP HISTORY FILE. 10/22/2023 00:00:00.000 DAILYTOTALSG670 (63). DAILYTOTALCOLLECTIONSTARTTIME: 10/21/2023 00:00:00.000. DAILYTOTALOFALLINSULINDELIVERED: 368250 (36.825 U). DAILYTOTALOFBASALINSULINDELIVERED: 212750 (21.275 U). DAILYTOTALOFBOLUSINSULINDELIVERED: 155500 (15.55 U). THERE WERE NO AUTOSUSPEND (12) ALARM NOTED IN THE PUMP HISTORY FILE. PLEASE SEE BELOW FOR THE USERSUSPENDED (2) ALARM LISTED ON THE EVENT DATE 21-OCT-2023 IN THE PUMP HISTORY FILE. 10/21/2023 16:01:05.000 INSULINDELIVERYSTOPPED (30). REASONFOINSULINDELIVERYSUSPENSION: USERSUSPENDED (2). THERE WERE NO UNEXPECTED PUMP ERROR(S)/ALARM(S) NOTED 1 WEEK PRIOR TO THE EVENT DATE 21-OCT-2023 IN THE PUMP HISTORY FILE. THE PUMP PASSED THE FUNCTIONAL TESTING. UNABLE TO CONFIRM ALLEGED HIGH BGS. COSMETIC DAMAGE WAS CONFIRMED AT THE BACK OF THE PUMP. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER EXPERIENCED HYPERGLYCEMIA WITH A BLOOD GLUCOSE VALUE OF 165 MG/DL AT THE TIME OF THE EVENT. THE CUSTOMER WAS TREATED IN AN HOSPITAL EMERGENCY ROOM WITH MANUAL INSULIN INJECTION AND THE CUSTOMER EXPERIENCED NO OTHER SYMPTOMS. TROUBLESHOOTING WAS PERFORMED AND FOUND THAT THE INSULIN PUMP HAD A CRACK ON THE BATTERY COMPARTMENT. IT WAS UNKNOWN WHETHER THE CUSTOMER WAS USING INSULIN PUMP WITHIN 48 HOURS PRIOR TO EVENT AND UNKNOWN WHETHER THE AUTO MODE FEATURE WAS ACTIVE AT THE TIME OF THE EVENT. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED. THE CUSTOMER WILL DISCONTINUE THE USE OF THE INSULIN PUMP AND IT WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1396482 MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1880 NG3195504H 000000763000545604

Patients

Seq Age Sex Outcome Treatment
1 48 YR Female Required Intervention FRN-MMT-332-RSVR,UNOMED INF SET