FDA Adverse Event Malfunction Summary report: N

TRUE METRIX AIR

MDR report key: 18088374 · Received November 7, 2023

Report

Report Number
1000113657-2023-00541
Event Type
Malfunction
Date Received
November 7, 2023
Date of Event
October 10, 2023
Report Date
November 30, 2023
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K150052
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

SECTIONS WITH ADDITIONAL INFORMATION AS OF 30-NOV-2023: H6: UPDATED FDA'S TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, AND INVESTIGATION CONCLUSIONS H10: METER AND. STRIPS WERE NOT RETURNED FOR EVALUATION. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIPS TESTED WITHIN SPECIFICATIONS. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-020: USER'S TEST STRIP HAD POOR STORAGE.

Additional Manufacturer Narrative · 0

INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-020: USER'S TEST STRIP HAD POOR STORAGE. NOTE: MANUFACTURER MADE SEVERAL ATTEMPTS TO CONTACT CUSTOMER TO ENSURE THE REPLACEMENT PRODUCT RESOLVED THE CUSTOMER¿S INITIAL CONCERN- UNABLE TO ESTABLISH CONTACT WITH CUSTOMER.

Description of Event or Problem · 0

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 402 AND 179 MG/DL. THE CUSTOMER¿S EXPECTED AM FASTING BLOOD GLUCOSE TEST RESULT RANGE IS P140-165 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION (KITCHEN). THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 03/31/2024 AND OPEN VIAL DATE WAS NOT DISCLOSED. THE CUSTOMER DID NOT HAVE ANOTHER VIAL OF TEST STRIPS THAT HAD BEEN STORED AND HANDLED CORRECTLY. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1: 159 MG/DL DATE: (B)(6) 2023 TIME: AM FASTING. RESULT 2: 133 MG/DL DATE: (B)(6) 2023 TIME: AM FASTING. RESULT 3: 121 MG/DL DATE: (B)(6) 2023 TIME: AM FASTING . RESULT 4: 402 MG/DL DATE: (B)(6) 2023 TIME: AM FASTING. RESULT 5: 179 MG/DL DATE: (B)(6) 2023 TIME: AM FASTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
922804 TRUE METRIX AIR SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, TMX HMN 50CT24/CASE MG/DL ZA4923S 00021292007836

Patients

Seq Age Sex Outcome Treatment
1 Unknown