Narratives

Additional Manufacturer Narrative · 0

CHECKING THE STERILIZATION CHARTS OF THE INVOLVED LOT #S, NO PRE-EXISTING ANOMALY WAS FOUND ON THE DEVICES. THEREFORE, PRODUCTS WITH THOSE LOT #S HAVE BEEN PROPERLY STERILIZED BEFORE BEING PLACED ON THE MARKET. THE ITEM INVOLVED WAS NOT AVAILABLE TO BE RETURNED TO LIMA-CORPORATE FOR FURTHER ANALYSIS. NO ADDITIONAL DETAILS WERE AVAILABLE ON THIS POST-OPERATIVE ISSUE, SPECIFICALLY PRE-OPERATIVE NOR POST-OPERATIVE X-RAYS RELATED TO THIS REVISION SURGERY WERE ACCESSIBLE. BASED ON THE VERY FEW INFORMATION RECEIVED, WE ARE NOT ABLE TO FURTHER INVESTIGATE THE ROOT CAUSE OF THE EVENT. HOWEVER, CONSIDERING THAT THE CHECK OF THE STERILIZATION CHARTS HIGHLIGHTED NO ANOMALIES ON THE COMPONENTS MANUFACTURED WITH THE INVOLVED LOT #S, WE CAN STATE THAT THE EVENT WAS NOT PRODUCT RELATED. PMS DATA ACCORDING TO LIMA-CORPORATE PMS DATA, REVISION RATE OF SMR REVERSE PROSTHESIS DUE TO INFECTION IS 0.079%. BASED ON THE ROOT CAUSE ANALYSIS PERFORMED AND ACCORDING TO THE RELEVANT PMS DATA, NO CORRECTIVE ACTIONS REQUIRED FOR THIS SPECIFIC CASE. LIMA-CORPORATE CONTINUES MONITORING THE MARKET TO PROMPTLY DETECT ANY SIMILAR ISSUE. NOTE: THIS IS A COMBINED INITIAL-FINAL MDR.

Description of Event or Problem · 0

SHOULDER REVISION SURGERY OF A SMR REVERSE PROSTHESIS PERFORMED ON (B)(6) 2021, DUE TO SUSPECTED INFECTION. THE FOLLOWING COMPONENTS GOT EXPLANTED AND REPLACED WITH NEW DEVICES: · SMR CEMENTLESS FINNED STEM (PRODUCT CODE 1304.15.160, LOT #1914651 - STER. 1900379). · SMR REVERSE FINNED HUMERAL BODY (PRODUCT CODE 1352.15.050, LOT #1907049 - STER. 1900191). · SMR REVERSE HP LINER MEDIUM (PRODUCT CODE 1365.09.015, LOT #1919049 - STER. 1900463) - PRODUCT NOT SOLD IN THE US. · SMR CONNECTOR SMALL R (PRODUCT CODE 1374.15.305, LOT #1920749 - STER. 2000011). · SMR REVERSE HP GLENOSPHERE 40 MM (PRODUCT CODE 1374.50.400, LOT #2003718 - STER. 2000072) - PRODUCT NOT SOLD IN THE US. THE PATIENT WAS THEN REVISED ON (B)(6), 2023, DUE TO IMPLANT LOOSENING (THE EVENT WAS REGISTERED AS COMPLAINTS #359_23 AND REPORTED TO THE FDA BY MFR 3008021110-2023-00123). PRIMARY SURGERY TOOK PLACE ON (B)(6) 2020. PATIENT IS A FEMALE, 69 YEARS OLD. EVENT HAPPENED IN AUSTRALIA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1283895 SMR CEMENTLESS FINNED STEM FINNED HUMERAL STEMS - 80MM (TI6AL4V) DIA.16MM HSD LIMACORPORATE S.P.A. 1304.15.160 1914651

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other