FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG SELF TEST 2CT

MDR report key: 18083760 · Received November 7, 2023

Report

Report Number
1221359-2023-01634
Event Type
Malfunction
Date Received
November 7, 2023
Date of Event
October 13, 2023
Report Date
November 7, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

FDA UDI: (B)(4). TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 210248 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160 / LOT 210248 AND TEST BASE PART NUMBER 195-430H / LOT 204649. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 210248 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE, INTERPRETATION OF THE RESULT, OR THE SPECIFIC PATIENT SAMPLE.

Description of Event or Problem · 0

THE CONSUMER REPORTED A FALSE POSITIVE RESULT WITH A BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON (B)(6) 2023 ON A NASAL SAMPLE. REPEAT TESTING WAS NOT PERFORMED. CONFIRMATION TESTING WAS NOT PERFORMED. CONSUMER PERFORMED AN ANTIGEN SELF-TEST WITH A DIFFERENT BRAND ON (B)(6) 2023 AND GENERATED A NEGATIVE RESULT. THE CONSUMER WAS ASYMPTOMATIC. THE CONSUMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THEIR TEST RESULTS. ADDITIONALLY, THE CONSUMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THEIR TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1174801 BINAXNOW COVID-19 AG SELF TEST 2CT CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 210248 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 73 YR Male