FDA Adverse Event Injury Summary report: N

NV GXL LINER NEUTRAL, 40MM ID, GROUP 4 CUPS

MDR report key: 18082751 · Received November 7, 2023

Report

Report Number
1038671-2023-02693
Event Type
Injury
Date Received
November 7, 2023
Date of Event
May 15, 2023
Report Date
April 24, 2026
Manufacturer
EXACTECH, INC.
Product Code
JDI
UDI-DI
10885862207104
PMA / PMN Number
K173583
Removal / Correction Number
Z-1732-2022
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H10. UPDATED/ADDITIONAL INFORMATION - H6. INVESTIGATION RESULTS - THE MOST LIKELY CAUSE FOR THE REVISION REPORTED DUE TO PROSTHESIS WEAR IS A COMBINATION OF RISK FACTORS INCLUDING, USE ERROR, IMPLANT POSITIONING, IMPLANT SIZE SELECTION, AND PATIENT FACTORS MAY HAVE ALSO BEEN A CONTRIBUTING FACTOR TO THE EARLY PROSTHESIS WEAR. HOWEVER, THIS CANNOT BE NOT CONFIRMED WITH THE INFORMATION PROVIDED; DEVICES WERE NOT RETURNED. THERE IS NO OTHER INFORMATION AVAILABLE.

Additional Manufacturer Narrative · 0

D10: CONCOMITANTS: (B)(6), 170-40-03 - BIOLOX DELTA FEMORAL HEAD 40MM 0D, +3.5MM. (B)(6), 186-01-60 - INTEGRIP CC, CLUSTER 60MM, G4. (B)(6), 188-01-07 - WEDGE PLASMA X/O SZ 7. PENDING INVESTIGATION.

Description of Event or Problem · 0

AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A RIGHT HIP REPLACEMENT ON (B)(6) 2016. APPROXIMATELY 6 YEARS AND 9 MONTHS AFTER THE INITIAL PROCEDURE THE PATIENT HAD A RIGHT HIP REVISION ON (B)(6) 2023. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN A COORDINATED ACTION IN (B)(6) WITH MASTER CASE NO. (B)(4). THE CONSOLIDATED LONG FORM COMPLAINT THAT APPLIES TO CASES FILED IN THIS COORDINATED ACTION ALLEGES THAT PATIENTS FILING SUITS IN THIS COORDINATED ACTION WERE REQUIRED ¿TO UNDERGO REVISION SURGERIES DUE TO SEVERE, PAIN, SWELLING, AND INSTABILITY¿ DUE TO ¿WEAR OF THE POLYETHYLENE COMPONENTS AND RESULTING COMPONENT LOOSENING AND/OR OTHER FAILURE FAILURES CAUSING SERIOUS COMPLICATIONS INCLUDING TISSUE DAMAGE, OSTEOLYSIS, PERMANENT BONE LOSS, AND OTHER INJURIES.¿ BECAUSE THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN THIS COORDINATED ACTION, THE PATIENT APPEARS TO ALLEGE THAT THE PATIENT WAS INJURED AS A RESULT OF WEAR OF AN EXACTECH POLYETHYLENE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
96287 NV GXL LINER NEUTRAL, 40MM ID, GROUP 4 CUPS PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED JDI EXACTECH, INC. UNK 10885862207104

Patients

Seq Age Sex Outcome Treatment
1