FDA Adverse Event Malfunction Summary report: N

BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE

MDR report key: 18076656 · Received November 6, 2023

Report

Report Number
9617032-2023-01590
Event Type
Malfunction
Date Received
November 6, 2023
Date of Event
October 20, 2023
Report Date
November 29, 2023
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
PMA / PMN Number
K022426
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: MATERIAL #: 364314. LOT/BATCH #: 2102900. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND FUNCTIONAL TESTING, EACH DRAWN WITH WATER, AND NO ISSUES WERE OBSERVED RELATING TO LEAKAGE AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE LEAKAGE. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE SMALL AMOUNT OF BLOOD WAS COMING OUT OF SYRINGE WALL AFTER DRAWING BLOOD GAS FOR A PATIENT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: A SMALL AMOUNT OF BLOOD WAS FOUND TO BE COMING OUT OF THE SYRINGE WALL AFTER DRAWING BLOOD GAS FOR A PATIENT, AND WAS REPLACED AND USED NORMALLY.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE SMALL AMOUNT OF BLOOD WAS COMING OUT OF SYRINGE WALL AFTER DRAWING BLOOD GAS FOR A PATIENT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: A SMALL AMOUNT OF BLOOD WAS FOUND TO BE COMING OUT OF THE SYRINGE WALL AFTER DRAWING BLOOD GAS FOR A PATIENT, AND WAS REPLACED AND USED NORMALLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338837 BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 2102900

Patients

Seq Age Sex Outcome Treatment
1 Unknown