FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 2.0 TEST KIT 24T OUS

MDR report key: 18076652 · Received November 6, 2023

Report

Report Number
1221359-2023-01628
Event Type
Malfunction
Date Received
November 6, 2023
Date of Event
October 4, 2023
Report Date
December 6, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GB
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

D4- UDI:(B)(4). THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 193-000 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 192-000. TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M233909 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 193-000/ LOT: M233909, TEST BASE PART NUMBER 192-430/ LOT: M233909. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M233909 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. H3 OTHER TEXT : SINGLE USE; DEVICE DISCARDED

Additional Manufacturer Narrative · 0

UDI:(B)(4). THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 193-000 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 192-000. THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. H3 OTHER TEXT : SINGLE USE; DEVICE DISCARDED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED ELEVEN (11) FALSE POSITIVE RESULTS AND ONE (1) FALSE NEGATIVE RESULT WITH THE ID NOW COVID-19 2.0 ASSAY PERFORMED ON THREE (3) DIFFERENT LOTS BETWEEN 01OCT2023 AND 22OCT2023 ON TWELVE (12) PATIENTS. THIS MFR. REPORT ADDRESSES LOT M233909 (TOTAL QUANTITY: (B)(4) AND PATIENT EIGHT (8) OF TWELVE (12). THE CUSTOMER REPORTED A FALSE NEGATIVE RESULT WITH THE ID NOW COVID-19 2.0 ASSAY PERFORMED ON 04OCT2023 ON A NASAL SAMPLE. CONFIRMATION TESTING VIA PCR (PLATFORM: PANTHER) WAS PERFORMED ON 05OCT2023 AND GENERATED A POSITIVE RESULT ON AN UNKNOWN SAMPLE TYPE ELUTED IN VTM. REPEAT TESTING WAS NOT PERFORMED. ALTHOUGH REQUESTED, NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED ELEVEN (11) FALSE POSITIVE RESULTS AND ONE (1) FALSE NEGATIVE RESULT WITH THE ID NOW COVID-19 2.0 ASSAY PERFORMED ON THREE (3) DIFFERENT LOTS BETWEEN (B)(6) 2023 ON TWELVE (12) PATIENTS. THIS MFR. REPORT ADDRESSES LOT M233909 (TOTAL QUANTITY: 5) AND PATIENT EIGHT (8) OF TWELVE (12). THE CUSTOMER REPORTED A FALSE NEGATIVE RESULT WITH THE ID NOW COVID-19 2.0 ASSAY PERFORMED ON (B)(6) 2023 ON A NASAL SAMPLE. CONFIRMATION TESTING VIA PCR (PLATFORM: PANTHER) WAS PERFORMED ON (B)(6) 2023 AND GENERATED A POSITIVE RESULT ON AN UNKNOWN SAMPLE TYPE ELUTED IN VTM. REPEAT TESTING WAS NOT PERFORMED. ALTHOUGH REQUESTED, NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338833 ID NOW COVID-19 2.0 TEST KIT 24T OUS REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. M233909

Patients

Seq Age Sex Outcome Treatment
1 Unknown