FDA Adverse Event Malfunction Summary report: N

SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

MDR report key: 18075448 · Received November 6, 2023

Report

Report Number
2249723-2023-04753
Event Type
Malfunction
Date Received
November 6, 2023
Report Date
July 11, 2025
Manufacturer
DATASCOPE CORP. - MAHWAH
Product Code
DSP
PMA / PMN Number
K112372
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 0

A SPARE HELIUM REGULATOR WAS ORDERED BY THE GETINGE FIELD SERVICE ENGINEER(FSE) THAT DISCOVERED THE ISSUE, BUT WAS FAULTY UPON INSTALLATION. UPON SUCCESSFUL INSTALLATION OF A WORKING REGULATOR, THE UNIT PASSED ALL SAFETY AND FUNCTIONAL TESTS AND WAS RETURNED TO THE CUSTOMER FOR CLINICAL USE.

Additional Manufacturer Narrative · 0

UPDATED FIELDS: B1, B4, D8, D9 (RETURN TO MANUFACTURER DATE), G1, G3, G6, H2, H3, H11. THE FOLLOWING INVESTIGATION WAS PERFORMED BY CARL FAMULARE, TECHNICIAN OF THE MAQUET FAILURE ANALYSIS AND TESTING DEPT. (FAT) WAYNE, NJ CF 21 OCT 2024. THE FAILURE ANALYSIS AND TESTING DEPT. RECEIVED PART NUMBER 0103-00-0637 HELIUM REGULATOR SERIAL NUMBER (B)(6) WITH A REPORTED UNIT FAILURE OF THE HELIUM BOTTLE COULD NOT BE USED. THE FAILURE ANALYSIS AND TESTING DEPT. PERFORMED A VISUAL INSPECTION AND FOUND THE PART TO BE IN GOOD CONDITION. THE FAILURE ANALYSIS AND TESTING DEPT. INSTALLED A KNOWN GOOD HELIUM TANK INTO PART NUMBER 0103-00-0637 HELIUM REGULATOR SERIAL NUMBER (B)(6) AND OBSERVED THAT THE HELIUM TANK COULD BE INSERTED INTO THE HELIUM REGULATOR WITH NO PROBLEM FOUND, PER THE CARDIOSAVE SERVICE MANUAL PART NUMBER 0070-00-0639-REV. R. THE FAT DEPT. COULD NOT REPLICATE THE FAILURE. THE HELIUM REGULATOR PASSED TESTING. RETAINING THE HELIUM REGULATOR IN THE FAT DEPT. PER PROCEDURE NUMBER 0002-07-D008 REV. AR. THE MOST PROBABLE ROOT CAUSE IS WITH THE YOKE OF THE HELIUM TANK NOT LINING UP WITH THE HOLES ON THE HELIUM REGULATOR. BR 21 OCT 2024. WHEN THE HELIUM TANK IS INSTALLED IN THE CARDIOSAVE IABP IT CONNECTS TO THE HIGH PRESSURE HELIUM REGULATOR WHICH IS LOCATED IN THE HOSPITAL CART. THE REGULATOR HAS TWO INDEXED PINS WHICH ALIGN WITH TWO HOLES ON THE HELIUM TANK VALVE. BOTH THE PINS ON THE REGULATOR AND THE HOLES IN THE TANK VALVE HAVE AN ALLOWABLE TOLERANCE FOR THEIR PLACEMENT. IF BOTH PARTS ARE WITHIN THEIR ALLOWABLE TOLERANCE, THEY WILL BE ABLE TO CONNECT. HOWEVER, A NUMBER OF 0202-00-0104 (ALSO REFERRED TO AS D202-00-0104) DISPOSABLE HELIUM TANKS HAVE BEEN IDENTIFIED AS HAVING THE LOCATION OF THEIR INDEX PIN HOLES OUT OF TOLERANCE. THIS OUT OF TOLERANCE CONDITION PREVENTS THE HELIUM TANK FROM CONNECTING TO SOME REGULATORS. ADDITIONALLY, THERE WERE A FEW REGULATORS THAT HAD PINS OUT OF TOLERANCE THAT WERE STILL ABLE TO BE SECURED INTO THE HELIUM TANK. THE ISSUES WITH THE HELIUM CYLINDER TANK ARE BEING INVESTIGATED UNDER 23-SCAR-CAHW-021.

Additional Manufacturer Narrative · 0

CORRECTED FIELDS: D9 (RETURN TO MANUFACTURER), H6 (TYPE OF INVESTIGATION). UPDATED FIELDS: B4, G3, G6, H2, H11.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE BEING PERFORMED BY A GETINGE SERVICE TERRITORY MANAGER (STM), THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) UNITS HELIUM BOTTLE CAN NOT BE USED WITH SPARE PART. THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 0

N/A.

Description of Event or Problem · 0

N/A.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1961939 SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. - MAHWAH

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown N/A.