SYRINGE 20ML LL TIP CONV PAK
Report
- Report Number
- 9610847-2023-00293
- Event Type
- Malfunction
- Date Received
- November 6, 2023
- Date of Event
- October 17, 2023
- Report Date
- January 23, 2024
- Manufacturer
- BECTON DICKINSON
- Product Code
- FMF
- UDI-DI
- 30382903056171
- PMA / PMN Number
- K110771
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER (B)(4) AND LOT NUMBERS 3212582, 3179209, 3118884, 3212582, 3088057, 3150615, AND 3150613. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. SHOULD YOU AGAIN EXPERIENCE ANY PROBLEMS WITH OUR PRODUCT WE WOULD APPRECIATE THE OPPORTUNITY TO CONDUCT A THOROUGH ANALYSIS. THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT THIS TYPE OF DEFECT DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. H3 OTHER TEXT : SEE NARRATIVE BELOW.
(B)(4) 1: INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. B3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. PATIENT PROBLEM CODE: F27 ¿ NO PATIENT INVOLVEMENT. DEVICE PROBLEM CODE: A180104 - DEVICE CONTAMINATION WITH CHEMICAL OR OTHER MATERIAL. LOT MANUFACTURE DATES: 3212582 - 2023-08-08 . 3179209 - 2023-07-27. 3118884 - 2023-05-18 . 3212582 - 2023-08-08 . 3088057 - 2023-05-02 . 3150615 - 2023-06-12 . 3150613 - 2023-06-05.
MAT#:(B)(4) BATCH#:3212582, 3179209, 3118884, 3212582, 3088057, 3150615, 3150613 IT WAS REPORTED BY THE CUSTOMER THAT THEY HAVE NOTICED THAT THERE ARE A LOT OF PARTICLES IN THE FINAL PRODUCT WHEN USING THE 20 CC SYRINGE TRAYS (ITEM # 305617). THEY HAVE ALREADY RULED OUT THAT IT IS NOT FROM THE DRUG. IT HAS BEEN SEEN IN EVERY LOT THEY CURRENTLY HAVE IN STOCK (3212582, 3179209, 3118884, 3212582, 3088057, 3150615, 3150613). VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. WE HAVE NOTICED THAT THERE ARE A LOT OF PARTICLES IN THE FINAL PRODUCT WHEN USING THE 20 CC SYRINGE TRAYS (ITEM # 305617). WE HAVE ALREADY RULED OUT THAT IT IS NOT FROM THE DRUG. IT HAS BEEN SEEN IN EVERY LOT WE CURRENTLY HAVE IN STOCK (3212582, 3179209, 3118884, 3212582, 3088057, 3150615, 3150613). CAN YOU PLEASE LOOK INTO THIS FOR US AS WE HAD TO DISCARD ALOT OF THE FINAL PRODUCTS AS WE CAN¿T SELL THEM WITH THIS ISSUE. 19 OCT 23 HELLO, THANK YOU FOR GETTING BACK TO US. HERE IS THE INFORMATION THAT YOU ASKED FOR. COULD YOU PLEASE PROVIDE OCCURRENCE DATE FOR THE INCIDENT?- WE HAVE SEEN THIS OCCURRING STARTING IN SEPTEMBER UNTIL RECENTLY. HOW MANY QUANTITIES AFFECTED IN THIS INCIDENT? IT WILL BE HELPFUL IF YOU CAN PROVIDE THE QUANTITIES FOR EACH LOT THAT HAS BEEN AFFECTED. 3212582, 3212582, 3088057, 3150615, 3150613- WE HAVE NO DATA ON THESE LOTS 3179209 (1045), 3118884 (1525), 3150617 (1579) COULD YOU PLEASE PROVIDE THE LOCATION WHERE THE PARTICLES HAVE BEEN FOUND?- THE PARTICLES WERE SEEN WHILE WE WERE VISUALIZING THE FINAL PRODUCT. WE SWIRL AND INVERT THE SYRINGE AND THE PARTICLE WOULD EITHER FALL DOWN OR RISE UP. IS THERE A PHOTO OR SAMPLE AVAILABLE FOR OUR REVIEW? IF SAMPLE IS AVAILABLE, PLEASE SEND ONLY 3-5 AFFECTED SAMPLES AND CONFIRM YOUR ADDRESS FOR RETURN LABEL. I WILL LOOK INTO THIS AND GET BACK TO YOU ABOUT THE SAMPLES.
MAT#:305617. BATCH#:3212582, 3179209, 3118884, 3212582, 3088057, 3150615, 3150613. IT WAS REPORTED BY THE CUSTOMER THAT THEY HAVE NOTICED THAT THERE ARE A LOT OF PARTICLES IN THE FINAL PRODUCT WHEN USING THE 20 CC SYRINGE TRAYS (ITEM # 305617). THEY HAVE ALREADY RULED OUT THAT IT IS NOT FROM THE DRUG. IT HAS BEEN SEEN IN EVERY LOT THEY CURRENTLY HAVE IN STOCK (3212582, 3179209, 3118884, 3212582, 3088057, 3150615, 3150613). VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. WE HAVE NOTICED THAT THERE ARE A LOT OF PARTICLES IN THE FINAL PRODUCT WHEN USING THE 20 CC SYRINGE TRAYS (ITEM # 305617). WE HAVE ALREADY RULED OUT THAT IT IS NOT FROM THE DRUG. IT HAS BEEN SEEN IN EVERY LOT WE CURRENTLY HAVE IN STOCK (3212582, 3179209, 3118884, 3212582, 3088057, 3150615, 3150613). CAN YOU PLEASE LOOK INTO THIS FOR US AS WE HAD TO DISCARD ALOT OF THE FINAL PRODUCTS AS WE CAN¿T SELL THEM WITH THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1917138 | SYRINGE 20ML LL TIP CONV PAK | SYRINGE, PISTON | FMF | BECTON DICKINSON | 3150613 | 30382903056171 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |