FDA Adverse Event
Injury
Summary report: N
SICAT CLASSICGUIDE
MDR report key: 18074037
·
Received November 6, 2023
Report
- Report Number
- 3006098230-2023-00001
- Event Type
- Injury
- Date Received
- November 6, 2023
- Date of Event
- October 18, 2023
- Report Date
- November 6, 2023
- Manufacturer
- SICAT GMBH & CO. KG
- Product Code
- NDP
- PMA / PMN Number
- NANA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
THE MOST LIKELY CAUSE OF THE DEVIATION IS THAT THE GUIDE WAS NOT HELD FIRMLY IN PLACE DURING SURGERY AND CHANGED POSITION WHEN FORCES WERE APPLIED BY THE DRILL.
Description of Event or Problem · 0
THE REPORTING DENTIST HAS USED A SICAT SURGICAL GUIDE (SICAT CLASSICGUIDE) FOR PREPARING OSTEOTOMIES (DRILL HOLES FOR ACCOMMODATING A DENTAL IMPLANT) FOR DENTAL IMPLANTS. HOWEVER, THE IMPLANTS AT TOOTH FDI #11, #15, #21, #25 DID NOT END UP AT THE CORRECT POSITION AND ORIENTATION. THEY ENDED UP TOO FAR IN THE DORSAL DIRECTION. AT POSITION #25 THE SINUS FLOOR WAS INJURED. THE DENTIST REMOVED THE IMPLANT #25 AND GRAFTED THE SITE AND WILL LET IT HEAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2131529 | SICAT CLASSICGUIDE | SURGICAL GUIDE | NDP | SICAT GMBH & CO. KG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Male | Required Intervention | CAMLOG GUIDE CA SL P+ |