FDA Adverse Event Injury Summary report: N

SICAT CLASSICGUIDE

MDR report key: 18074037 · Received November 6, 2023

Report

Report Number
3006098230-2023-00001
Event Type
Injury
Date Received
November 6, 2023
Date of Event
October 18, 2023
Report Date
November 6, 2023
Manufacturer
SICAT GMBH & CO. KG
Product Code
NDP
PMA / PMN Number
NANA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE MOST LIKELY CAUSE OF THE DEVIATION IS THAT THE GUIDE WAS NOT HELD FIRMLY IN PLACE DURING SURGERY AND CHANGED POSITION WHEN FORCES WERE APPLIED BY THE DRILL.

Description of Event or Problem · 0

THE REPORTING DENTIST HAS USED A SICAT SURGICAL GUIDE (SICAT CLASSICGUIDE) FOR PREPARING OSTEOTOMIES (DRILL HOLES FOR ACCOMMODATING A DENTAL IMPLANT) FOR DENTAL IMPLANTS. HOWEVER, THE IMPLANTS AT TOOTH FDI #11, #15, #21, #25 DID NOT END UP AT THE CORRECT POSITION AND ORIENTATION. THEY ENDED UP TOO FAR IN THE DORSAL DIRECTION. AT POSITION #25 THE SINUS FLOOR WAS INJURED. THE DENTIST REMOVED THE IMPLANT #25 AND GRAFTED THE SITE AND WILL LET IT HEAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2131529 SICAT CLASSICGUIDE SURGICAL GUIDE NDP SICAT GMBH & CO. KG

Patients

Seq Age Sex Outcome Treatment
1 69 YR Male Required Intervention CAMLOG GUIDE CA SL P+