MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL
Report
- Report Number
- 2032227-2023-297548
- Event Type
- Injury
- Date Received
- November 6, 2023
- Date of Event
- September 12, 2023
- Report Date
- November 6, 2023
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000414344
- PMA / PMN Number
- P150001
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- 003
Narratives
(B)(4). THE PUMP PASSED THE SELF TEST, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST AND DAT AT 0.08725 INCHES. HOWEVER, THE PUMP HAD A HIGH SLEEP CURRENT MEASUREMENT. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUMP. SUCCESSFULLY UPLOADED PUMP TO CARELINK. IN FURTHER FULL REVIEW OF THE PUMP HISTORY/TRACES ON THE SS EVENT DATE OF (B)(6) 2023, THERE IS NO UNEXPECTED ALARMS/SUSPENDS AND NO BOLUS DELIVERY NOTED. PLEASE SEE BELOW FOR THE DAILY TOTAL OF BASAL/BOLUS AND ALL INSULIN DELIVERED ON THE SS EVENT DATE (B)(6) 2023 LISTED ON SMARTSOLVE. DAILY TOTAL OF ALL INSULIN DELIVERED: 231500 (23.15 U). DAILY TOTAL OF BASAL INSULIN DELIVERED: 231500 (23.15 U). DAILY TOTAL OF BOLUS INSULIN DELIVERED: 0 (0 U). THE PUMP WAS PROGRAMMED WITH MULTIPLE BOLUS DELIVERIES AND ALL BOLUS DELIVERED PROPERLY THEIR INDICATED AMOUNTS (AT QUICK BOLUS SPEED) AND WERE PROPERLY RECORDED IN THE DAILY HISTORY. NO BOLUS DELIVERY ANOMALY OR HISTORY ANOMALY NOTED. THERE WERE NO UNEXPECTED PUMP ERRORS/ALARMS NOTED 1 WEEK PRIOR TO THE EVENT DATE (B)(6) 2023 IN THE FORMATTED HISTORY FILE. POWER MANAGEMENT GRAPH WAS SUCCESSFULLY GENERATED. THE POWER MANAGEMENT TOOL CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. NO POWER ERROR 25, LOW BATTERY ALERT, POWER LOSS ALARM, REPLACE BATTERY ALERT OR REPLACE BATTERY NOW ALARM WERE FOUND IN THE HISTORY FILES OR TRACES ON THE EVENT DATE. THE PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND NO EVIDENCE OF PHYSICAL OR MOISTURE DAMAGE ON THE PCBA 1, PCBA 2, FORCE SENSOR, MOTOR AND VIBRATOR ASSEMBLY NOTED. THE ORIGINAL PCBA 2 WAS INSTALLED IN A TEST PCBA 1, CASE, INTERNAL BATTERY AND MOTOR. POWERED THE PUMP ON USING THE BATTERY SIMULATOR AND CONTINUED TESTING. THE PUMP FAILED THE SLEEP CURRENT MEASUREMENT. THE ORIGINAL PCBA 1 WAS INSTALLED IN A TEST PCBA 2, CASE, INTERNAL BATTERY AND MOTOR. POWERED THE PUMP ON USING THE BATTERY SIMULATOR AND CONTINUED TESTING. THE PUMP FAILED THE SLEEP CURRENT MEASUREMENT. A HIGH SLEEP CURRENT MEASUREMENT (UNEXPECTED BATTERY POWER LOSS) WAS CONFIRMED, SUSPECTED ON THE PCBA 1/PCBA 2. FORCE SENSOR ZERO OFFSET WITHIN SPECIFICATION (23.6 MV). THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE ON THE NGP STB3 AND PASSED. THE PUMP WAS RECEIVED WITH THE ORIGINAL BATTERY CAP. NO CRACKED/DAMAGED BATTERY CAP NOTED DURING VISUAL INSPECTION. THE PUMP WAS RECEIVED WITH A 1.419V ENERGIZER MAX ALKALINE BATTERY INSTALLED. THE TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A CRACKED KEYPAD OVERLAY, A STAINED KEYPAD OVERLAY AND A SCRATCHED CASE. UNABLE TO VERIFY CUSTOMER ALLEGED FOR HIGH BGS/DKA. THE FORCE SENSOR IS WITHIN SPECIFICATION AND THE MOTOR FUNCTIONING PROPERLY. HOWEVER, A HIGH SLEEP CURRENT MEASUREMENT (UNEXPECTED BATTERY POWER LOSS) WAS CONFIRMED, SUSPECTED ON THE PCBA 1/PCBA 2. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER EXPERIENCED HYPERGLYCEMIA WITH A BLOOD GLUCOSE VALUE OF 225 MG/DL AT THE TIME OF THE EVENT. THE CUSTOMER FELL UNCONSCIOUS AND WAS HOSPITALIZED. THE CUSTOMER DID NOT EXPERIENCE ANY SYMPTOMS RELATED TO HIGH BLOOD GLUCOSE. THE CUSTOMER WAS TESTED POSITIVE FOR KETONES. TROUBLESHOOTING WAS PARTIALLY PERFORMED. THE CUSTOMER HAD USED THE INSULIN PUMP SYSTEM WITHIN 48 HOURS OF THE REPORTED HIGH BLOOD GLUCOSE EVENT. THE CUSTOMER USED THE SMARTGUARD AUTO MODE OF THE INSULIN PUMP. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED. THE CUSTOMER HAD DISCONTINUED THE USE OF THE INSULIN PUMP AND WAS RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1960661 | MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1880 | HG5BHJMZZ | 000000763000414344 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR | Female | Required Intervention| O| H | FRN-MMT-332A-RSVR, UNOMED INF SET. |